VELOCITY ACCURACY 

 

Set flow pump so that the average flow velocity is 20 cm/s. See “Pump Calibration” (page 11) for 

instructions on converting volume flow rate to average flow velocity.

 

 

To ensure laminar flow, the scan region should be placed outside the minimum entrance length (>6 cm at 

maximum flow velocity).

 

 

Enable Pulse-Wave mode. Image the tube in cross-section, set angle correction to that vessel angle and 

set the pulse wave window to encompass the entire lumen. Estimate the maximum velocity (center of the 

tube). For laminar flow, the maximum velocity will be double the average flow velocity.

 

 

Enable Color Flow mode. Scan flow tube in the cross-sectional position. Set the pulse repetition frequency 

(PRF) to avoid aliasing and decrease until aliasing sets in. The maximum velocity corresponding to the PRF 

setting is the maximum velocity in the flow tube.

 

Note: Color Flow only yields an average velocity, which should match the calibrated average flow velocity of 

the flow pump.

 

 

 

ANGLE ERROR TEST 

1. 

Set the flow circuit for constant velocity flow. 

 

Enable Color Flow mode. Scan the flow tube in the longitudinal sectional position.

 

 

3 . 

For orientation 1, which tests for abdominal applications, the angle between the membrane and the vessel 

is 45°. Thus, the Doppler angle is 45° if the probe is placed normal to the membrane. The transducer should 

be placed closer to the shallow end to measure the angle at the middle region of the tubing.

 

 

4 . 

For orientation 2, which tests for peripheral applications, the angle between the membrane and the vessel is 

20°. The Doppler angle is 70° if the probe is placed normal to the membrane.

 

 

5 . 

Position the Doppler window on the center of the tube and correct the angle to that of the vessel. Record the 

observed angle and freeze the image.

 

 

6 . 

Measure the angle between the membrane and beam direction by using the caliper application. The corrected 

Doppler angle is given by the equation given below:

 

 

Содержание 069A

Страница 1: ...USER GUIDE 900 Asbury Ave Norfolk Virginia 23513 USA Tel 757 855 2765 WWW CIRSINC COM Model 069A...

Страница 2: ......

Страница 3: ...structions for Use 4 Handling and Care 4 Testing Procedures 7 Sensitivity Depth of Penetration 7 Velocity Accuracy 9 Angle Error Test 9 Specifications 10 ZERDINE 11 Appendix 1 Conversion of Flow Rate...

Страница 4: ...edescribedinthis manual whiletheReferencessectionprovidessourcesforadditional information about Doppler Ultrasound Performance Testing Description of the Phantom Flowphantomcontainsablood vessel simul...

Страница 5: ...with the following acoustic properties Speed of sound 1540 m s Attenuation coefficient 0 7 dB cm MHz Relative backscatter 0 dB with respect to CIRS liver parenchyma reference HANDLING PRECAUTIONS AND...

Страница 6: ...s original weight and you notice a difference in vertical distance measurements or the scan surface appears depressed call CIRS at 800 617 1177 This product contains Zerdine a non flowing water based...

Страница 7: ...rom the pump and leave the tube in the air 2 Run the pump pumping until all remaining fluid is returned to the reservoir 3 Turn off the pump disconnect the reservoir cap and place it over a container...

Страница 8: ...ty based on the system settings and operator technique In addition there may be some measurement bias between the pump settings and the velocity measurement There fore it is recommended that when usin...

Страница 9: ...l view of the vessel is recommended for the most consistent results Move the transducer down the length of the scanning surface so that the vessel depth increases Check for flow periodically until flo...

Страница 10: ...e Color Flow only yields an average velocity which should match the calibrated average flow velocity of the flow pump ANGLE ERROR TEST 1 Set the flow circuit for constant velocity flow 2 Enable Color...

Страница 11: ...ansducers it will be necessary to measure the probe angle for each measurement An example of a Doppler angle measurement with a curved transducer is shown in the image below 7 Compare the observed ang...

Страница 12: ...Hz Scatter or relative contrast in the optimal range of 15 to 15 dB in relation to a scatter baseline equivalent to human liver tissue Elasticity with a Young Modulus in the optimal range of 4 to 90 k...

Страница 13: ...you are not satisfied with your purchase for any reason please contact Customer Service or your local distributor prior to returning the product Visit https www cirsinc com distributors to find your...

Страница 14: ...ouble the average flow velocity In pulsatile mode the maximum flow velocity depends on the shape of the pulse profile Assuming a simple sine wave the peak flow velocity will be double the maximum flow...

Страница 15: ......

Страница 16: ......

Отзывы: