MK:427 Environmental Noise Microphone
Page 13 of 23
Example
Actuation output measured = 89.9 dB
Actuation level on Factory Configuration Sheet = 90.6 dB
So, difference between output and factory tested setting = 0.7dB or 0.7/10 = -0.07mA
Typical time history trace during using the electrostatic actuation process
The picture below shows the effect of the actuation process upon the output level produced by the
MK:427.