5.4-
Test cell extraction
If still it has not been done, open the door of the equipment and pull out the test cell from the
chamber. In order to extract the test cell, pull out both of white supports simultaneously and firmly,
one with each hand.
5.5-
Adjustment of the distance between electrodes
In order to adjust the distance between electrodes we will use the gauge delivered with the
equipment. The gauge must be placed between the electrodes, and the distance between them
will be corrected relaxing its subjection by means of provided Allen spanner. Later it will be
necessary to tighten to maintain the required distance.
Remarks:
The supplied gauge follows the IEC-156 standard. This Standard specifies that the
gauge must be 2,5 mm ± 0,05 mm in width. If other standards are required, you should
use a different gauge under request (Go to 10
th
point)