Operation Manual

                                

   

NORM-O-TEMP, Model 111W

  

 
 

Page 4 of 26 

Cincinnati Sub-Zero Products, Inc., reserves the right to make changes to the device, which may not be 
reflected in this manual. 

WARNING 

 

A physician's order is required for the use of the device and setting the temperature of the 

blanket/pad. At least every 20 minutes, or as directed by physician, check patient's temperature 
and skin condition of areas in contact with blanket/pad; also, check blanket/pad water 

temperature. Pediatric patients, temperature-sensitive patients with vascular disease, surgical 
patients, diabetic and Reynaud’s disease patients should be checked more frequently.  If patient’s 

temperature does not reach desired set point or differs drastically from recommended set point, 
notify physician. 

Notify the physician promptly of any change in patient status in order to 

avoid serious injury or death.

  

1.

 

Pediatrics – Infant’s and children’s body temperatures are often more responsive to 
surface heating and cooling than adults.  Due to their size, the effect of heating or 
cooling a child is likely more pronounced because of their higher ratio of skin contact 

area to body mass. 

2.

 

Temperature Sensitive Patients – Patients with impaired peripheral blood circulation 
and patients who are incapacitated may be more sensitive to temperature changes 

than patients with normal circulation. 

3.

 

Surgical Patients – Patients with poor circulation due to inadequate heart function, loss 
of blood, or impaired peripheral blood circulation may be more sensitive to 

temperature changes. 

 

The method of temperature control provided by all hyperthermia units presents the danger of 

heating body tissues, particularly the skin, to a point where they are injured. Depending on the 
extent and severity of a burn, 

very serious and even fatal complications may arise.  

 

 

Do not use the NORM-O-TEMP

®

 System distal to arterial cross clamping. 

Thermal injury may 

result.

 

 

Prevent excessive and/or prolonged tissue pressure and shearing forces, especially over bony 

prominences, to prevent 

skin damage that may result.

 

 

Do not place additional heat sources between the patient and blanket/pad. 

Skin damage may 

result.

 

 

The area between the patient and the blanket/pad should be kept dry to avoid injury to 

patient. 

 Prep solutions have been reported to injure the skin when allowed to remain between 

patients and a water-circulating heating blanket/pad during prolonged procedures.  

 

Proper sanitation procedures must be practiced and hygienic safety must be maintained, 
to prevent contamination.

  Contamination can affect patient’s health, i.e. skin irritation/rash 

may result.

 

 

Do not use the NORM-O-TEMP

®

 system in the presence of flammable anesthetics. 

Risk of 

explosion can result.

 

 

Power interruption will cause the NORM-O-TEMP

®

 unit to revert to Preset Set point Temperature 

resulting  in  possible  inaccurate  therapy  to  the  patient.  Follow  instructions  for  First  Time  Set-

Up/System  Test  Routine  located  in  the  Operation/Technical  Manual  to  resume  operation.  

Failure to resume therapy could result in serious injury or death. 

 

Do  not by-pass ground lug.

   Electrical Hazards may result.

 

 

Any time water is found leaking into or around the unit, connecting hose, and/or blanket/pad, 

turn the unit off, disconnect the power cord from its power source, and correct the problem 
before  proceeding.

  Water  leaks  could  lead  to  electric  shock.    Water  leaks  could  also 

present a slip hazard.  

 

Содержание NORM-O-TEMP

Страница 1: ...Operation Manual Model 111W Hyperthermia System Cincinnati Sub Zero Products Inc 12011 Mosteller Road Cincinnati Ohio 45241 U S A www cszmedical com ...

Страница 2: ... 26 NORM O TEMP MAXI THERM MAXI THERM LITE Gelli Roll and PLASTI PAD are registered trademarks of Cincinnati Sub Zero Products Inc Cincinnati Ohio USA Copyright 2015 Cincinnati Sub Zero Products Inc All rights reserved Manual 57127 Rev O ECN M409 4826 ...

Страница 3: ...g Temperature Decreasing Temperature Disconnect Power Before Servicing Water Flow Indicator Outlet Inlet Silence Alarm Caution Read Operation Instructions and Manual Before Operating Temp Set Point Indicator Fill to Strainer Fuse Heat Separate collection for electrical and electronic equipment Protective Earth Ground Equipotentiality Earth Ground Clean Water Filter Quarterly Caution Low Water Leve...

Страница 4: ...thermia units presents the danger of heating body tissues particularly the skin to a point where they are injured Depending on the extent and severity of a burn very serious and even fatal complications may arise Do not use the NORM O TEMP System distal to arterial cross clamping Thermal injury may result Prevent excessive and or prolonged tissue pressure and shearing forces especially over bony p...

Страница 5: ...ays be performed Improper repair and inadequate maintenance can result in damage to the NORM O TEMP system and patient injury Remove the NORM O TEMP unit from service if the outer casing or membrane control panel is cracked or internal components are exposed Contact with internal components could result in electric shock or thermal injury to the patient or operator and exposure to sharp edges Keep...

Страница 6: ...per Electrostatic Discharge ESD practices should be followed during replacement of any electronic board To Discharge the NORM O TEMP system ensure the unit is unplugged and use any metal tool such as a screwdriver with an insulated handle to simultaneously touch both pins on the power cord The operator must regularly monitor the patient whenever hyperthermia or normothermia therapy is used For saf...

Страница 7: ...1 3 5 External Features and Descriptions Top View 15 1 4 Required Accessories 17 Section 2 Specifications and Certifications 17 2 0 Unit and Patient Related Precautions 19 2 1 Patient Preparation and Bedside Care 19 2 2 Alarms and Error Displays 19 Section 3 Operating the NORM O TEMP System 21 3 0 Introduction 21 3 1 Arranging the System Components 21 3 2 Operating the NORM O TEMP System 22 3 3 Co...

Страница 8: ...r of your NORM O TEMP unit ready when you call for parts or service The serial number is located on the specification label on the back of the unit In Warranty Repair and Parts All parts on your NORM O TEMP unit are covered by a one year 1 warranty Additional warranties are available at the time of purchase or during the warranty period To return defective parts or units first obtain a Returned Ma...

Страница 9: ...he set point temperature of 37 C Observe Water Flow Indicator on the side of the unit to verify that the pump is circulating water when blanket pad is connected The unit will start heating to a preset temperature of 37 C To change the desired set point temperature proceed to step 5 below 5 Depress the TEMP SET button The display will flash the previous set point temperature or the default temperat...

Страница 10: ...Alarm button The Silence Alarm button will not silence the alarm when ERR appears on the display or HIGH TEMP indicator light comes on and does not flash Power cord should be removed from outlet and the unit sent to Biomedical Engineering for repair CAUTION This device should be operated only by trained personnel under the direction of a physician Patients vary in degree of sensitivity to cold hea...

Страница 11: ...procedures and to reduce cold discomfort before during and after a surgical procedure The thermal regulating system is used to keep a patient comfortable by maintaining blanket pad water temperature through conductive heat transfer The water heated blankets transfer the thermal energy to adult pediatric and infant includes neonates patients to keep a patient at a comfortable temperature The NORM O...

Страница 12: ...tion Technical Manual B The operating instructions printed below the control panel describe the necessary steps to operate the unit C The power switch is a bevel rocker switch labeled I on at the top and O off at the bottom D The four rubber feet help the unit to mount on the I V stand Catalog 118 the low profile stand Catalog 119 or allow the unit to sit on a flat surface E The water fill opening...

Страница 13: ... the speed of the paddle wheel The water flow indicator only spins when a blanket pad or by pass hose is connected to the unit It will not spin when water is circulated internally to pre condition the water A total obstruction of the water path will cause the paddle wheel to stop completely B Two screws on the left and right side of the unit secures the top to the base C Two female Hansen fittings...

Страница 14: ...on for the internal components C The nylon straps are used to secure and store the coiled power cord and or the connecting hose when not in use D The serial number printed on the Spec Label and located just above the power cord E Units have a power cord that can be disconnected and should only be inserted into a properly grounded mating receptacle of a hospital grade specified by local codes and p...

Страница 15: ...he membrane control panel as shown in Figure 5 is composed of pressure sensitive touch switches and LED displays The membrane control panel is divided into the following sections A The four digit LED display shows the actual water temperature and set point temperature B When the HEAT LED is lighted it indicates that the water is being heated C The HI TEMP LED will illuminate the power fail LED wil...

Страница 16: ...n is used to display and change the desired set point temperature When this button is pressed the display will flash the set point temperature G The INCREMENT and DECREMENT buttons are used to raise or lower the set point temperature The operator has less than ten 10 seconds to set the temperature up or down after pressing the TEMP SET button H The fill reservoir lid slides to the left to open and...

Страница 17: ...lastic top Dual reservoir Built in handles Alarms High Temperature Audible and Visual Low Water Audible and Visual Defective Water Temperature Sensor Audible and Visual Water Flow Indicator Visual Power Failure Audible and visual Greater than 1 Over Set point Audible and Visual Control System Microprocessor based temperature control system and Alarm indications Controller Range Water Temperature H...

Страница 18: ...2 C Primary Temperature Limit 43 5 C 0 6 C Secondary Temperature Limit 44 5 C 0 6 C Independent Mechanical Temperature Limit 46 C 0 6 C Environmental Conditions during storage and transportation Ambient Temperature transportation and storage 40 C 50 C Humidity transportation and storage 20 95 Warranty 1 year parts Parts and labor if returned to factory Additional Warranty Available UL Electrical C...

Страница 19: ...ian if the patient s core temperature does not reach the prescribed temperature in the time prescribed or deviates from the prescribed temperature range B Changes in skin color edema inflammation or indications of pressure especially over bony prominences should be noted and treated as ordered Avoid prolonged tissue pressure and shearing forces over bony prominences C The patient should be turned ...

Страница 20: ... high temperature safety will activate to shut off the power to the pump shut off the power to the heater the HI TEMP LED and Power Fail LED illuminate and the trouble alarm sounds 2 If the independent mechanical high temperature alarm occurs the unit should be powered off and the power cord removed from the outlet The unit should be immediately removed from service and sent to Biomedical Engineer...

Страница 21: ...needed carefully add distilled water Do not use de ionized water Do not overfill If the water falls below a preset level the alarm sounds and the LO WATER LED lights up The pump and heater shut down and the operator cannot proceed until this is corrected by refilling the unit and resetting the alarm F Inspect the power plug for bent or missing prongs Do not by pass ground lug Electrical hazards ma...

Страница 22: ...ctuating the power switch I O switch a power fail alarm will activate and the Power fail symbol LED located on the front of the unit will flash To stop the alarm restore power 3 2 Operating the NORM O TEMP System The NORM O TEMP system operates based upon the actual temperature of the circulating water relative to the set point temperature The NORM O TEMP system heats and circulates the water Give...

Страница 23: ...arterly Dispose of the NORM O TEMP unit per hospital protocol 4 1 Replenishing the Reservoir Fixing a Low Water Alarm A Check that the drain hose is disconnected B Slide the water fill opening lid to the left and gradually pour approximately 1 gallons 5 7 liters of distilled water with approximately 340 cc of propylene glycol into the reservoir DO NOT USE ALCOHOL TAP WATER OR DE IONIZED WATER C St...

Страница 24: ... environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power supply lines 1 kV for input output lin...

Страница 25: ...strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected b...

Страница 26: ... Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated us...

Страница 27: ......

Страница 28: ...Cincinnati Sub Zero Products Inc 12011 Mosteller Road Cincinnati OH 45241 Telephone 1 800 989 7373 or 513 772 8810 Fax 513 772 9119 E Mail cszinc cszinc com ...

Отзывы: