STACK-O!
Testing System
CEP STACK-O! Manual Rev. 2.1
User’s Manual
July 15, 2016
Page 5
The SENSOR SETUP screen allows the user to calibrate the STACK-O!, and enable and disable
the sensors. (By default, both sensors are enabled. A sensor should only be disabled when the
sensor has failed. If the user disables one of the sensors, the user should recalibrate the STACK-
O!.
) Press “Calibrate Sensors” The following screen will be displayed:
Figure 5
– STACK-O! SENSOR CALIBRATION Screen
The SENSOR CALIBRATION screen, when used in conjunction with the STACK-O! 10-inch
calibration rod supplied with the STACK-O!, allows the user to calibrate the STACK-O!. The
calibration procedure is as follows:
1. Remove all product from the STACK-O! Product Holder and make sure that the Product
Holder is clear of all debris. If the STACK-O! is equipped with the Tare Plate Option,
place the Tare Plate on the Product Holder.
2.
Press the “Set 0”” button on the screen. This sets the 0” point in the STACK-O!’s
controller.
3.
Place the 10” calibration rod on the Product Holder. If the STACK-O! is equipped with
the Tare Plate Option, place the Tare Plate on top of the 10” calibration rod.
4.
Press the “Set 10”” button on the screen. This sets the 10” point in the STACK-O!’s
controller.
5. The STACK-
O! is now calibrated. Press the “Return to Setup” button to return to the
SENSOR SETUP screen and then press the “Setup Done” button to exit the SENSOR
SETUP screen and to display OPERATOR measurement screen.
Remove the 10”
calibration rod from the Product Holder. If the STACK-O! is equipped with the Tare Plate
Option, allow the Tare Plate to hang by its tether outside the Product Holder.
You can restart the calibration procedure at any time by pressing the “Restart Cal. Proc.” Button.
NOTE
Use of the Tare Plate is optional. The Tare Plate is meant to be used with product material that has
an uneven surface. If the user decides to start or stop using the Tare Plate after initial calibration, the
STACK-O! should be recalibrated accordingly.