2.4.
Eye Safety
The CE376 incorporates a laser which is rated as a class IV device. This is an embedded laser
where the output from the sensor unit, through the optics, is minimized to class 1M. This
classification indicates that viewing of the beam with the naked eye is safe but looking directly
into the beam with optical instruments, e.g. binoculars can be dangerous.
From the laser head the output has the following characteristics:
•
Wavelength:
532 nm / 808 nm
•
Maximum pulse energy:
7 µJ / 4 µJ
•
Pulse duration:
1 ns / 150 ns
•
Pulse frequency:
4.6 kHz
•
EN 60825-1:2001
The laser output may vary with time as well as the repetition rate. In all case, the standard
LiDAR emission will be eyesafe.