(5) “[TIME]” Button (System time setting) 

To set or change the system time, press the [TIME] 
button in the setup mode. The time and date on the 
third display panel should blink. Enter from left to 
right in the following format YY:MM:DD and 
HH:MM:SS. Press number button printed on the over-
lay in white word to set the system date and time.   
Press the “ENTER” button to confirm. Exit this func-
tion by pressing ESC button. 

Note:

 When it shows low battery “

”and when after 

changing battery will reset the date and time to default 
setting. To ensure that the setting is correct, please 
change new battery and reset the date and time manu-
ally when “

”appears on the LCD display. 

 

PROCEDURE OF CALIBRATION 

1. Make ADJ pin short. 
2. T2 input a voltage 70mV, it will show a value higher 

than 28000 ±2dgts at the third display and then press 
“set[ ]” key into next step. 

3. T1 input a temperature 0°C, it will show a value ±5°C at 

the main display and then press “set[ ]” key to complete 
the calibration. 

4. Take off the short pin.

 

 

Error messages: 

When the meter appears wrong messages such as Err-01, 
Err-02 and Err-03. 
Err-01: In the “SAVE” mode, the maximum recording 

ranges are 128 samples. The lower right display 
will show Err-01 if the maximum ranges are over 
128 samples. 

Err-02: In the “LOG” mode, the maximum recording are 

16*64=1024 samples. The lower right display will 
show Err-02 if the maximum ranges are over 16 
groups. 

Err-03: In the “READ LOG” and “READ SAVE” mode, 

when the main display shows 6208 and the lower 
right display shows “OL, Err-03”. Which means 
that there might appear two situations: 

1. There is no data in the memory. 
2. The memory is full, and the meter will warn the user 

this is the last sample. 

 SAVE=128 

samples 

  LOG= 1024 samples 

OPERATOR MAINTENANCE 

WARNING 

To avoid possible electrical shock, disconnect the 

thermocouple connectors from the thermometer 

before removing the cover.

 

 

Battery Replacement 

Power is supplied by a 9 volt “transistor” battery. (NEDA 
1604, IEC 6F22). The “

” appears on the LCD display 

when replacement is needed. To replace the battery, re-
move the two screws from the back of the meter and lift 
off the battery cover. Remove the battery from battery 
contacts. 
 

Cleaning 

Periodically wipe the case with a damp cloth and detergent, 
do not use abrasives or solvents. 
 

 

Ver. 1.5 05/10/2021

17008-502A1-2