Freeze-dryer Epsilon 2-6D LSCplus
8 Maintenance and service
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
123 / 156
Translation of the original operating manual
• The other oil changes depend on the operating conditions. In general,
an interval of 500 to 1,000 operating hours is sufficient.
• Please contact the Christ service department if you have any queries
(see chapter 7.3 - "Service contact").
Pos : 258 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 259 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/080 W artung und I nst andhaltung/ 080-0010- 0083 Vak uumsy st em Epsil on (Tr ock enl äufer pum pen) LSCpl us, LPCpl us @ 45 \mod_1475738107196_68. doc x @ 443597 @ 3 @ 1
→
8.1.8
Vacuum system with a dry-running vacuum pump
The dry-running vacuum pump requires only minimal maintenance.
NOTE
Comply with the maintenance instructions of the manufacturer in the
separate operating instructions of the vacuum pump and silencer!
In addition, the following must be taken into consideration:
WARNING
Risk of burns on hot surfaces
During the operation of the freeze-dryer and half an hour afterwards, the
outer surface of the vacuum pump may be hot.
There is a risk of burns when touching the surfaces.
• Prior to performing any maintenance work on the vacuum pump, let
the components cool!
• Wear suitable protective clothes and gloves!
WARNING
Danger of poisoning/infection due to contaminated exhaust air
The exhaust air may contain harmful substances originating from the
product.
Contact with particles in the exhaust air (especially by inhalation) may
result in severe damage to health.
• Use a silencer with integrated filter elements or – depending on the
product
– guide the exhaust gases from the outlet of the vacuum pump
to a suitable treatment system in order to prevent the discharge of
harmful substances into the surrounding atmosphere! Compliance with
the national rules and regulations for the protection of the environment
must be ensured!
• When working on the vacuum system (and especially on the vacuum
pump), it may be necessary to wear suitable protective clothing,
gloves and a breathing mask depending on the product!
The number of operating hours of the vacuum pump can be monitored by
way of the LSCplus control system and also by way of the CHRIST SCADA
software LPCplus (see chapter 6.6.1.3 - "Main window
"Options""/"Operating hours" and "
Vacuum pump oil change” or separate
operating manual of the Christ SCADA software LPCplus, index entries
"Working hours" and "Oil change interval (vacuum pump)").