Christ Elektronik CAM 100 Скачать руководство пользователя страница 2

 

1 Description 

1.1 General 

With the digital panel meter CAM 100 temperatures are measurable with Pt100 sensor.  The connection of the Pt100 sensor 
can take place in 2-, 3- or 4-wire technology.  Two different measuring range can be programmed.  Despite the small dimen-
sions the LED display can be read off even from larger distance well. 

1.2 Safety instructions 

This equipment is built and according to quality standards examined in accordance with European guidelines.  It left the work 
in safety-relevant perfect condition.  The references and warning notes contained in this operating instructions must be con-
sidered around a safe enterprise to ensure.  Without impairment of its working reliability the equipment can be operated 
within the certified site conditions.  See chapter 3.  This equipment may be taken only by a specialist in enterprise, which is 
familiar with the associated dangers and/or the relevant regulations. 

1.3 Maintenance 
All repairs of the device may only be carried out by a specialist workshop. In case it is inevitable to carry out repairs on the 
opened device which is still supply voltage this may only be effected by a trained specialist who knows about the dangers 
usually related to any such procedure. In case of misuse or wrong operation of the decice we do not assume any liability for 
any damages that might occur. 

1.4 Mounting 
The equipment is to be inserted from the front into the cutout planned for it (according to DIN 43 700).  Dimensions of the 
cutout:  45 x 22.2 [mm].  The attachment takes place with the help of enclosed fastening parts.  The tightening screws are to 
be tightened mutually, until the equipment sticks.  During the placement of the equipment is the radiant heat of neighbouring 
devices to consider (consider permissible ambient temperature!).  The electrical connection is to be made according to appro-
priate regulations (e.g. VDE 0100).  Supply voltage is indicated on the type plate and is to put on the clamps 5 and 6. 

 

2. Operation 

2.1 Starting 

With the digital panel meter CAM 100 the temperature - measuring range can be stopped by programming places.  The deci-
mal point is likewise set with programming places. 

2.2 Adjustment Pt 100  
1.    Select the temperature display range (see table 2) by closing the solder pads (see picture 2). 
2.    Connect the Pt100 Simulator to terminal block corresponding to table 1 / picture 1. 
3.    Bring the display to the required value for span start using the zero potentiometer “NP” (see pict. 1). 
4.    Apply the value for span end with the Pt100-simulator to terminal block. 
5.    Bring the display to the required value for span end using the gain potentiometer „V“ (see pict. 1). 
6.    Repeat step 3 to 5 until the display corresponds to the required measuring range. 

 

 
2.3 Rear view and solder pads 

Picture 1:    Connecting diagram 

 

 

 

 

 

 

 

 

 

 

 
 
 
 
 
 
 

 

 

 

1  2  3  4  5  6 

                                         B1      B2 

Bridges: 
2-wire B1 and B2 
3-wire B1 

          Pt100 

4-wire no bridges 

+   - 

5, 12, 24 V

DC 

Supply voltage

 

Gain Potentio-
meter „V“ 

Zero Potentio- 
meter „NP“ 

Содержание CAM 100

Страница 1: ...lpenstrasse 34 D 87700 Memmingen Tel 49 0 83 31 83 71 0 Fax 49 0 83 31 83 71 99 About this manual Read this manual carefully for safety instructions and operating guidance before installation Operation manual CAM 100 Digital panel meter for Thermometer Pt100 ...

Страница 2: ...The equipment is to be inserted from the front into the cutout planned for it according to DIN 43 700 Dimensions of the cutout 45 x 22 2 mm The attachment takes place with the help of enclosed fastening parts The tightening screws are to be tightened mutually until the equipment sticks During the placement of the equipment is the radiant heat of neighbouring devices to consider consider permissibl...

Страница 3: ...r Pads closed 2 4 wire 100 0 199 9 C 2 1 2 4 wire 200 700 C 1 and 4 3 wire 100 0 199 9 C 3 3 wire 200 700 C 1 and 4 1 Preset delivery configuration Table 1 Connecting Pt100 Connection terminal block bridges 4 wire to terminal block 1 2 3 4 no bridges 3 wire to terminal block 1 3 4 and bridge B1 2 wire to terminal block 1 a 4 and bridge B1 a B2 DP 1 DP 0 DP 2 DP 1 DP 3 DP 3 DP 2 DP 0 ...

Страница 4: ...4 wire sensors each max 100 Ω resistance Wire effect 3 wire connection 0 01 K Ω 4 wire connection 0 002 K Ω Distribution 4 wire connection Meas range 100 0 199 9 C Supply voltage Supply volltage 5VDC 4 5 5 5VDC 12VDC 12 13 2VDC 24VDC 22 3 26 4VDC galvanically separated ripple max 100 mVSS Protection Protection against pole reversal EMC References According to European Directive 89 336 EWG Electrom...