5. Safety Warnings
DANGER
•
Do not use this device if you have an implanted defibrillator or implanted metallic
devices, such as a pacemaker. Such use could cause electrical shock, burns, electrical
interference or death.
•
Don't use the device if you use a high frequency surgical ME EQUIPMENT
simultaneously. Such use may result in burns and possible damage to the device.
•
Operation in close proximity (e.g. 1m) to a shortwave or microwave therapy ME
EQUIPMENT may produce instability in the device output.
•
The application of electrodes near the thorax may increase the risk of cardiac fibrillation.
WARNING
•
Consult your physician before purchasing or using this device if you have one of the
following conditions:
-Acute disease
-Malignant tumor
-Infective disease (i.e. HIV/AIDS, Pneumonia)
-Pregnant
-Heart disease
-High fever
-Abnormal blood pressure
-Lack of skin sensation
-Abnormal skin condition
-Any condition requiring the active supervision of a physician
•
Use of accessories and cables other than those specified or provided by the manufacturer
of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
•
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the
ME EQUIPMENT, including cables specified by the manufacturer. Otherwise, degradation
of the performance of this equipment could result.
•
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.