Please check the batteries voltage.
b) "SEn oFF " alarm
Please check if the probe was connected with the oximeter
correctly. If the sensor is with extension cable please check if the
extension cable is connected with the sensor correctly.
c) "E2" alarm
The signal is weak, please check the patient condition.
d) E1
Change the SPO2 module, please contact to the manufacture.
10.4 Warranty and Repair
10.4.1 Maintenance Method
a
)
Maintenance responding time: 9:00am
~
17:30pm, Monday to Friday
b
)
Service support: Our company will offer user telephone and e-mail
technology support and parts change.
Parts change: our company will change parts if it is necessary free
of charge in the warranty period.
Because parts are the sources of maintenance, user should send
them back to our company if not specified.
c
)
Update the system software free of charge.
10.4.2 Exempt and limitation:
a) Our company isn’t responsible for such damage caused by force
nature. For example: fire, thunder flash, flood, cyclone, hail,
earthquake, house collapse, commotion, plane failing and traffic
accident, deliberate damage, lack of fuel or water, labor and capital
bother, strike and stop-working etc.
b) No-service offer
The corresponding charge and insurance charge of disassembling,
refurbishing, repackaging and moving the oximeter or the part of it.
The damage caused by the third company not commended by our
company to adjust, install replace the parts of the oximeter.
The damage and failure caused by user or its representative doesn’t
comply with the operator’s manual
c) The oximeter is installed or connected with such external device
without our company permission as printer, computer, internet line
and lead to oximeter failure. Our company will charge for the
maintenance.