INST.No.INE-508-P0
-1-
1.1 General
The IR-E series scanning radiation thermometer, having a rotary mirror built in the detector
unit, is designed to measure unidirectional temperature patterns of an object without contact
and performs the temperature measurement of 50°C to 1200°C.
The IR-E series consists of the detector unit (Model IR-E) and the high-speed
display/processing unit (Model IR-EPG), and exclusive accessories of the protective case and
the air-purge hood for the detector unit are added as needed.
The detector unit measures the thermal radiation energy from a measuring area by scanning
measuring positions optically by the rotary mirror, and converts the measured temperature into
an electric signal that is transmitted to the high-speed display/processing unit with an
exclusive connecting cable.
The connecting cable is shipped with the length specified, but the cable length can be changed
within 200m without any effect to functions.
The high-speed display/processing unit performs indication/recording and analog output of the
measured temperature after arithmetic processing such as emissivity compensation,
linearization, signal modulation, etc. on the temperature pattern signal transmitted from the
detector unit.
2 kinds of connection types of a connector connection and a terminal connection are available
on the detector unit. The connection type of your detector unit has been selected by your
ordering specification.
This manual describes operations and configurations of the detector unit, the protective case
(Model IR-ZEPW), and the air-purge hood (Model IR-ZEAH). Confirm the name and model of
the instrument you purchased and read corresponding paragraphs in this manual.
A separate manual is prepared for handling/operation of the high-speed display/processing unit,
read it along with this manual.
1.2 Model code
Detector unit
IR-E
(a×b)
Distance factor 0 : 50 ( 50× 50)
1
:
100
(100×100)
2
:
200
(200×100)
Connection type
T : Terminal connection
C : Connector connection
1.3 Measurement ranges
50 ~ 300°C, 100 ~400°C, 200 ~ 600°C, 300 ~ 800°C, 400 ~ 1000°C, 500 ~ 1100°C,
600 ~ 1200°C
1. Introduction
The term of the distance
factor indicates one side of
the scanning direction. For
the instantaneous view,
refer to the following
figure.
Distance factor a (200)
Distance factor b (100)
Scanning direction
Instantaneous
view
Ì
The term of the distance factor is the
“distance factor a”.
The numbers shown in parentheses
are for the distance factor of 200.
C aution