INST.No.INE-559-P1
-6-
5. Connections and Wirings
5.1 Connection to the Main unit
Detector is connected prior to a shipment. Connect a cable at the power supply and signal line. Use
dedicated cable or commercial cable (twisted-pair cable with a shield, outer diameter
φ
6.5 to
φ
8.0,
wire size 0.2 to 0.5mm2.). Connect cable shield to FG. If using thermocouple output, use extension wire
according to the thermocouple type. Conditions for cable length are as follow.
Cable length
Wire size
Power supply voltage
5m or shorter
0.2mm
2
(AWG24) or more
8V or higher
50m or shorter
0.2mm
2
(AWG24) or more
9.5V or higher
100m or shorter
0.3mm
2
(AWG22) or more
9.5V or higher
200m or shorter
0.3mm
2
(AWG22) or more
11V or higher
(1) Loosen four of M3 screws of Main unit and remove a lid.
(2) Set a cable to the cable gland and connect to a terminal block. Press a button on the terminal block
while inserting the wire and release the button after the insertion.
To prevent an electric shock, turn OFF the power source for wiring.
M3 screw
W a rn i ng
!
Terminal block
Terminal block
Cable gland
Button
Wire
9
If taking out the lid, please conduct at clean
condition such as no dust.
If take out the lid, please do not touch the
electronical parts other than terminal block
by screw driver or remove the screw.
W a rn i ng
W a rn i ng
F
G
8
-
3
6
V
D
C
0
V
D
C
S
A
S
B
D
I
D
O
C
O
M
A
G
N
D
O
U
T
-
A
m
A
O
U
T
-
T
C
O
U
T
-
H
T
Содержание IR-BZ Series
Страница 2: ......
Страница 11: ...INST No INE 559 P1 5 4 3 Measuring Diameter and Measuring Distance Unit mm φ16 φ8 400 200 ...
Страница 43: ......