- 111 -
12. Inspection and Maintenance
12-1. Routine Inspection
Check the remaining amount of chart and recording condition on a daily basis to keep the unit in good condition. When
any abnormality is found, take an appropriate action according to the “11. Troubleshooting”.
Maintenance/inspection item
Operation
Plotter pen and cartridge
pen replacement
When the trace printing (trend line) fade away or becomes less visible, replace
the cartridge pen with new one.
When the letters of digital printing fade away or become less visible, replace the
plotter pen with new one.
(See “6-1.2. How to set plotter pen and cartridge pen”.)
Chart replacement
A chart paper can last about a month when fed at a speed of 25mm/H
continuously.
When the end of chart nears, an end mark (red line on the right edge of
chart) appears. In this case, replace the chart with new one.
(See “6-1.1. How to set chart paper”.)
Cleaning
Wipe away dirt on the unit with a soft, dry cloth or a cloth dampened with
warm water or neutral detergent.
Do not use chemical solvents including thinner and benzine to prevent the unit surface
from melting. The chart guide is made of acrylic. It may produce a crack when a
chemical solvent is used.
Do not lubricate the mechanical section such as the main shaft and gear.
12-2. Consumable Parts and Replacement Guideline
The unit includes some consumable parts. To use the unit for a long time in good condition, we recommend that these
parts be replaced regularly.
Do not replace parts other than chart and pens by yourself. Not only does it fail to replace
properly, but it also may pose dangerous situation. Make sure to contact CHINO’s sales
agent for replacement of consumable parts.
1. Consumable parts and recommended replacement cycle
(Usage under the condition of temperature: 20 to 25°C, humidity: 20 to 80%RH, operation time: 8hours/ a day)
Consumable part
Replacement cycle
Other operating condition
Me
ch
an
ic
al
pa
rts
Pen servo (for cartridge pen)
4 to 6 years
Use under the normal condition as below:
No corrosive gas
Dust free, oil smoke free, dry place
Free from vibration and impact
No other factors affecting operation
Pen servo (for plotter pen)
4 to 6 years
Belt
4 to 6 years
Chart drive mechanism
4 to 6 years
Motors
4 to 6 years
Electron
ic p
art
s
Power supply
5 years
At an ambient temperature of 25°C
Relay (for alarm)
100,000 times
Resistive load
30,000 times
Inductive load
Lithium battery
10 years
8-hour operation per day (at an ambient
temperature of 40°C or lower)
Keys 500,000
times
Depends highly on the use and surrounding
conditions.
Display (LCD)
4 to 6 years
At an ambient temperature of 25°C
!
Caution
!
Warning
!
Caution