3.3 WELL MOUNTED DETAILS
Handrail adjustment
Gas spring
Gas spring of tilting kit
(only 01327)
Gas spring with
spacers
If after mounted the parallel bar doesn’t look like the details in the pictures please contact our
assistance service department.
4 ADJUSTMENTS
4.1 HANDRAIL HEIGHT ADJUSTMENT
1
Grasp the handrails, release the position by pressing the pedal control (same effect achieved
on right and left pedal)
2
Bring the handrails to the desired height and release the pedal to maintain the position.
WARNINGS!
During the handrail height adjustment, the springs alone will not ensure that the patient is secure on
the platform
4.2 HANDRAIL WIDTH ADJUSTMENT
Unscrew one of the two knobs, on each handrail, until removing the pin from the three-hole blade.
Loosen the other knob (on the same handrail) and pull the handgrip to remove the pin from the
three-hole blade; at this stage, while carrying out this operation with one hand, use the other hand
to grab the handrail, which will now be free to move and can therefore be adjusted in width, by
inserting the pin into the desired hole.
Tighten this knob and then screw back and tighten the other one.
The handrails, when tightened to the central hole, allow using the glider, code 01380