background image

Operating Manual:

CPF01L86

www.chinastirling.com

19

6. Maintenance & Service

The CPF01L86 Ultra-low Transportable Cryogenic Freezer is designed for years

of trouble-free operation without a major maintenance. To prevent costly and

inconvenient repairs and maintain your Freezer to an optimum level of

performance, follow the recommended preventative cleaning maintenance and

contact an authorized service provider if needed.

6.1. Cleaning the Freezer

Clean the exterior surfaces of the CPF01L86 Ultra-low Transportable Cryogenic

Freezer as needed by using a soft cloth and mild detergent, do not use solvent

(such as bleach) or harsh abrasive cleansers or pads.

When changing the goods or in the event of excess frost accumulation inside the

freezer, turn OFF and empty the Freezer compartment and clean it with soft dry

clothes and avoid using any sharp objects to remove the frost.

6.2. Air Intake Filter Cleaning

Clean accumulated dust and dirt on the air intake filter every month. Remove the

screws of side cover to open it and separate the filter from the unit. Gently clean

the filter with a vacuum cleaner. If there are stubborn residues use a soft brush

to work them loose.

6.3. Managing Freezer Contents

The CPF01L86 Ultra-low Freezer is designed for storage and transport of

different materials such as, Vaccine, biological samples and other medical

purpose usage which require ultra-low and well-regulated temperature.

The temperature set-point for the freezer should be changed as appropriate

for the materials being stored using the User Interface (UI) as described in

the section 5.

Materials may be placed in the freezer in any convenient arrangement that

does not block or prevent the door of freezer from closing completely.

For best results operate freezer at full capacity with real or simulated

product to increase the thermal mass and maintain optimal stability.

Disinfect with suitable sterilizing agent if the freezer has been used for

bio-hazards, before using it for any other goods or material.

Before putting the freezer in storage when not used, turn OFF the power

and allow the freezer to come to room temperature and dry the inside of the

freezer compartment and clean any spills.

Содержание CPF01L86

Страница 1: ...CPF01L86 Ultra low Transportable Cryogenic Freezer Operating Manual 1 Stirling road north economic develop zone rizhao city shandong province China www chinastirling com ...

Страница 2: ......

Страница 3: ...rigeration 5 3 2 Power and Temperature Control 5 3 3 Power Adapter and Cord 5 3 4 Thermal Insulation 5 3 5 Graphical User Interface 6 3 6 Tracking and Monitoring 6 4 Unpacking and Setup 9 4 1 Unpacking 9 4 2 Setup 9 5 Operation 11 5 1 Responsible Operation 11 5 2 UI operation guidelines 11 5 3 Error Alarms and Troubleshooting 17 6 Maintenance Service 19 6 1 Cleaning the Freezer 19 6 2 Air Intake F...

Страница 4: ...Operating Manual CPF01L86 www chinastirling com ...

Страница 5: ...re storage and portable carrying for different products with a stable AI controlled deeply frozen environment This product is Using a user friendly UI User Interface which is easy to understand and use It is classified for use as stationary equipment in a Pollution Degree 2 and Over voltage Category II environment The unit is designed to operate under the following environmental conditions Indoor ...

Страница 6: ...mperatures are dangerous use proper precautions when operating at ultra low temperatures China Stirling cannot be held responsible for damages or loss of stored product attributed to unintended use In no case will China Stirling be held liable for loss of stored product resulting from electrical mechanical or structural failure As with any ultra low temperature freezer appropriate back up and redu...

Страница 7: ...the unit alone Keep infant away from unit Risk of burn and electric shocks Do not touch inside the freezer with wet hands while operating on ultra low temperatures Risk of frostbite Do not touch any electrical parts such as the power supply plug or any switches with a wet hand That may cause electric shock lead to death or injury If there is an unexpected sound smell smoke when the power is turned...

Страница 8: ...d plug not cable There is Danger of causing electric shock or short circuit Avoid strong physical shock or dropping of objects on the unit This may cause damage to electronic components inside Remove the plug before turning off engine of the vehicle Power does not turn off after switching off the engine in some vehicles freezer can draw power from car battery Do not put ice or liquid water directl...

Страница 9: ...equipped with built in battery providing the backup power to freezer for one hour to give stable temperature all the time G pro control board with AI algorithm programming control the FPSC power to maintain the temperature of freezer temperature sensors PT100 measures the internal freezer temperature and machine cold head temperature and provides input to the controller The UI provides the display...

Страница 10: ...Monitoring Equipped with the new Stage mOS 3 0 and LOTA FPSC control hardware built in Glonass GPS global positioning and timing system 4G LTE Wi Fi Bluetooth Zigbee all weather stable connection remote start and stop configuration monitoring real time temperature curve 6DOF motion processor collects and analyzes all equipment motion data opening a new chapter for your precise medical products tra...

Страница 11: ...tems Lid latch for positive closure 3 Freezer lid Freezer is equipped with well designed insulated lid Providing backup power to freezer for 1 hour 4 Lid gasket Outer moisture seal 5 Freezer cabinet Metallic cabinet with proprietary coating and with double insulation for better protection of stored items 6 Air vent and Filter Intake air for cooling of FPSC and filter to protect it from dust 4 6 5 ...

Страница 12: ...th foam strong easy to carry 8 Power Switch and DC Socket Switch to turn ON OFF the power to Freezer DC socket is tagged with polarized signs to avoid errors 9 Backup Battery Replaceable battery with locked cover Providing backup power to freezer unit for more than 1 hour 10 Exhaust Vent ...

Страница 13: ...the freezer and all accessories for any shipping damage Check the packing list to verify that the shipment is complete Included Items Packing List Operating Manual this document Qualification Certificate Freezer AC Adapter Power Cord Cigarette lighter cable Optional Unpack freezer unit and included components ensure all packaging material has been removed around the air inlet and outlet holes 4 2 ...

Страница 14: ...ord to attach with power adapter Use the B type DC plug to connect the adaptor to Freezer insert the plug in to DC socket press the plug firmly into the connection port Turn ON the Freezer by using the switch on the machine above the socket Once the power switch is turned ON unit will be powered and display screen will be ON after a BEEP Follow the instructions from section 5 of this operating man...

Страница 15: ...il in the UI guideline section Operation for the AC and DC system is described in the section 4 operation in vehicle DC system need extra precautions to avoid any incidents Notes on operation in a vehicle 1 Connect cigarette lighter plug to vehicle after starting the engine 2 The vehicle engine should be running during the operation of Freezer to prevent accidental discharge of the battery 3 Do no...

Страница 16: ...ch screen Program version number cab be seen on the display screen Have a Quick View at Simple operation procedure programming is done for friendly interface and equipment running Keep on reading for details about the UI and it s usage The interface description The program has designed a total of 6 six operation interfaces the boot interface the main interface the Wi Fi setting interface and the W...

Страница 17: ... 3 Display the current mode off cooling heating 4 Display the set target temperature 5 Change target temperature 6 Temperature curve showing the current temperature curve 7 Display the current temperature inside the cabinet 8 Display ambient temperature 9 Machine running time 10 Press to lock or unlock the cabinet door 11 Press to enter the user setting interface 12 Press to enter Wifi configurati...

Страница 18: ...of 5 parts 1 Back button click to return to the previous interface 2 Official LOTA page QR code 1 Scan QR code 2 Pay attention to the public number 3 Follow the page prompts after scan 3 Wi Fi configuration 4 OTA upgrade 5 Program version number 4 Wi Fi configuration operation instructions This interface consists of 4 parts ...

Страница 19: ...setting interface operation instructions The interface consists of 11 parts 1 Back button click to return to the previous interface 2 Automatic switch button 3 Turn on the password button click to set the password according to personal needs ON OFF 4 RFID management NFC card management 5 Screen off time display click to set the screen display time 6 Brightness setting display click to set the disp...

Страница 20: ...er control temperature The machine contains 5 thermometers The temperature of T1 is used as the master control temperature 5 Reporting interval the time interval for reporting data to the server 6 Temperature compensation 1 T1 temperature compensation 7 Upper temperature limit upper target temperature setting 8 Temperature compensation 4 T4 temperature compensation 9 Lower temperature limit lower ...

Страница 21: ...t point Setting value is too small Modify alarm current setting Other Component damage Contact after sales to return to the factory for maintenance E3 High temperature at hot end Hot end temperature is more than 55 High ambient temperature ambient temperature Use in the working temperature range Air outlet blocked Air outlet blocked Keep the air outlet unobstructed Fan 1 does not turn Fan 1 not wi...

Страница 22: ... not wired Check the fan 2 wiring Fan 2 damaged Contact after sales to return to the factory for maintenance EB Fuse open DC voltage is 0V Fuse blown Fuse blown Replace fuse EC Insufficient power supply DC voltage less than 11 8V Battery powered Battery depletion Replace power supply or battery Power supply Power supply error Check or replace the power supply ...

Страница 23: ...f side cover to open it and separate the filter from the unit Gently clean the filter with a vacuum cleaner If there are stubborn residues use a soft brush to work them loose 6 3 Managing Freezer Contents The CPF01L86 Ultra low Freezer is designed for storage and transport of different materials such as Vaccine biological samples and other medical purpose usage which require ultra low and well reg...

Страница 24: ... the adapter tightly Assure that the wall plug is firmly seated DC plug is not properly inserted Check the DC plug and insert in the socket firmly Car cigarette lighter fuse has blown Change the fuse and try to start the Freezer Freezer does not achieve desired set point The door seal of the Freezer is not good Inspect the inner door and door seal for damage and replace if necessary Inadequate air...

Страница 25: ...l Conditions Non corrosive non flammable non explosive Indoor use Altitude up to 2 000m Temperature 20ºC to 45ºC Maximum relative humidity 80 for temperature up to 31 C decreasing linearly to 50 at 40 C Internal Volume 01L External Dimensions 270x311x450 mm L x W x D Internal Dimensions 150x134x54 5 mm L x W x D Net Weight 13 5kg Insulation High performance vacuum insulated panels VIP and polyuret...

Страница 26: ...ns and Description Dimensions of China Stirling Ultra low Transportable Cryogenic Freezer are shown in the pictures below all the dimensions are in mm millimeter Major internal parts of the unit is also clearly mentioned in the pictures different views Front View Side View ...

Страница 27: ...astirling com 23 Top View 8 3 Freezer Performance Characteristics Ultra low Transportable Cryogenic Freezer performance test graph is given below Freezer reach 86 from ambient 25 in less than two hours when freezer is in empty condition ...

Страница 28: ...na Stirling be responsible or held liable for consequential or incidental damages associated with loss of stored product in the event of an equipment failure Extended warranty programs are available Contact China Stirling for a custom warranty solution International Distributor Limited Warranty Warranty will start ONE MONTH after the ship date from China Stirling China Stirling warrants that Distr...

Страница 29: ......

Страница 30: ...oad North Economic Development Zone Rizhao city Shandong province China 86 15000777883 86 17763336212 www chinastirling com Specifications subject to change without notice Refer to www chinastirling com for the latest specifications ...

Отзывы: