Check-line AD-100 Скачать руководство пользователя страница 5

– 4 –

3.0   memory Pointer (optional)

The Memory Pointer moves 

with the measurement pointer 
when a measurement is taken. 
However, when pressure on the 
specimen is released, and the 
measurement pointer returns to 

zero, the Memory Pointer remains 

in position, marking the last 
measurement.

To return the Memory Pointer to 

the zero position prior to taking 
another measurement, rotate the 
knurled knob on the face of the 
instrument counter- clockwise until the pointer has been reset. 

mEmOry POINT-

Er

KNurlEd KNOb

Содержание AD-100

Страница 1: ...Operating Instructions CHECK LINE BY ELECTROMATIC Analog Durometer Model AD 100 ...

Страница 2: ... 1 0 Introduction 2 1 1 Unpacking 1 2 Complete Kit 2 0 Testing Procedure 3 3 0 Memory Pointer 4 4 0 Specifications 5 4 1 Conforms to International Standards 4 2 Dimensional Drawings 5 0 Calibration Repair 7 6 0 Warranty 8 ...

Страница 3: ... are available A B C D DO O OO The AD 100 conforms to ASTM D 2240 DIN 53505 ISO 868 ISO 7619 1 1 Unpacking Unpack the durometer and inspect it for any shipping damage Notices of defect must be filed immediately in writing at the latest within 10 days on receipt of the goods 1 2 Complete Kit The AD 100 is supplied as a complete kit including the durometer instruction manual and NIST traceable calib...

Страница 4: ...of the gauge firmly against the specimen but not so firmly as to imbed the foot into the surface of the material 3 Maintain pressure for 2 to 3 seconds The dial hand gives the reading in durometer points Important After the initial reading has been noted continue maintaining pressure for several more seconds Creep or cold flow of the specimen if present will be evident by receding action of the di...

Страница 5: ...imen is released and the measurement pointer returns to zero the Memory Pointer remains in position marking the last measurement To return the Memory Pointer to the zero position prior to taking another measurement rotate the knurled knob on the face of the instrument counter clockwise until the pointer has been reset Memory Point er Knurled Knob ...

Страница 6: ...se ranges be omitted Measuring Range 0 to 100 units Accuracy 1 durometer unit A B C D O DO 2 durometer units OO Dial Size 1 97 50mm diameter Dimensions 5 04 x 2 24 x 1 18 128 x 57 x 30 mm Weight 7 ounces 200 grams Warranty 1 year 4 1 Conformance to International Standards Model ASTM D2240 DIN 53505 ISO 868 ISO 7619 AD 100 A AD 100 B AD 100 C AD 100 D AD 100 DO AD 100 O AD 100 OO ...

Страница 7: ... 6 0 71 in 18mm 0 98 in 25mm 0 79 in 20mm 2 24 in 57mm 5 04 in 128mm 8oz 1 00 in 25 4mm 4 1 Dimensional Drawings ...

Страница 8: ... 7 5 0 Calibration and Repair All Check Line durometers requiring calibration and or certification should be sent directly to ELECTROMATIC Equipment Co Inc ...

Страница 9: ...E RESULTING FROM THE USE OF SUPPLIED EQUIPMENT WHETHER SUCH DAMAGE OCCURS OR IS DISCOVERED BEFORE UPON OR AFTER REPLACEMENT OR REPAIR AND WHETHER OR NOT SUCH DAMAGE IS CAUSED BY MANUFACTURER S OR SUPPLIER S NEGLIGENCE WITHIN ONE YEAR FROM INVOICE DATE Some State jurisdictions or States do not allow the exclusion or limitation of incidental or consequential damages so the above limitation may not a...

Страница 10: ...Distributed by ABQ Industrial LP USA Tel 1 281 516 9292 888 275 5772 eFax 1 866 234 0451 Web https www abqindustrial net E mail info abqindustrial net ...