Functional Description
68
Dual-Channel Digital Oscilloscopes
Functional Description
UTILITY System (cont’d)
RUN Pass/Fail test
Follow next steps
1. Press UTILITY button to enter the “Utility menu”.
2. Press “Next Page ” to enter the third page of “Utility menu”
3. Press “Pass/Fail” to enter the “Pass/Fail” function menu.
4. Press “Enable Test” to select “Open”.
5. Press “Source” to select input signal channel.
6. Press “Next Page” to enter the second page of the “Pass/Fail menu”.
7. Press the “Mask Setting” button to enter the first page of “Mask
menu”.
8. Press “X Mask”, turn the “Universal” knob to adjust the horizontal
clearance.
9. Press “Y Mask”, turn the “Universal” knob to adjust the vertical
clearance.
10. Press “Create Mask” to create mask and you can also can the next
page of “Mask menu” to recall the stored mask.
11. Enter the second page of “Pass/Fail function menu”, press “Output”
to set the output option.
12. Enter the first page of the “Pass/Fail function menu”, press
“Operate” to select “
” and to run the pass/fail test.
Note
The output signal is 3 V. When an error is detected, the output signal is
0 V during 125 µs.