Chauvin Arnoux C.A 757 Скачать руководство пользователя страница 10

21

2.3. RESISTANCE, CONTINUITY, DIODE AND 

CAPACITANCE

 

„

Connect the red probe tip to the 

+

 terminal and the black 

probe tip to the 

COM

 terminal. 

 

„

Set the switch to 

W

.

 

„

Keep your hands behind the guards of the device and of 

the test probe.

 

„

Place the test probes on the element to be tested (Figure 

12). If a voltage is present, the instrument so indicates.

 

Do not make a resistance, diode, continuity, or capac-

itance measurement on a live circuit.

As default, the instrument is in automatic mode (

SCAN

) and 

chooses automatically among the resistance 

, continui-

ty 

, diode 

, and capacitance 

 functions. To force 

one of these functions, press the 

Select

 key (the 

SCAN

 

symbol disappears).

Resistance and continuity

If the resistance is less than 300 

W

, the instrument is in con-

tinuity (Figure 11). Below 30 

W

, it emits a continuous beep.

Between 300 

W

 and 3 M

W

, the instrument is in resistance 

mode. Above 3 M

W

, the display unit indicates OL. The 30 M

W

 

range is not available in automatic mode; resistance mode 

 must be used.

 

Diode

 

„

In diode test, place the red probe tip on the anode of the 

diode to be tested and the black probe tip on the cathode.

 

„

The instrument indicates the voltage of the diode. If it 

is above 2V or if the polarity is reversed, the instrument 

displays - - - -.

Capacitance 

The 3mF and 30mF ranges are not available in automatic 

mode; capacitance mode 

 must be used.

If the device displays 

dis.C

 after a measurement, wait for 

the end of automatic discharging of the capacitance before 

making another measurement.

2.4. CURRENT

 

„

Connect the current sensor to the 

+ terminal

 

„

Set the switch to 

A

.

 

„

Press on the yellow opening device to open the flexible 

coil (Figure 13).

 

„

Open it and place it around the conductor carrying the 

current to be measured (Figure 14) (only one conductor 

in the sensor).

 

„

Close the coil. To optimize measurement quality, centre 

the conductor in the coil and make the coil as round as 

possible.

 

„

The current is displayed.

北京海洋兴业科技股份有限公司 电话:010-62176775

Содержание C.A 757

Страница 1: ...E Bedienungsanleitung IT Manuale d uso ES Manual de instrucciones Testeur num rique Digital tester Digitalen Pr fer Tester digitale Comprobador digital C A 757 600 V CAT III 1000 V max COM DIGITAL TES...

Страница 2: ...illustrazioni ilustraciones 1 2 3 4 5 6 7 8 9 COM Select R 10 11 12 15 SCAN Position limite des mains Position beyond which your hands must not go u erste Position der H nde Posizione limite delle ma...

Страница 3: ...emplacement des piles Les piles et les accumulateurs usag s ne doivent pas tre trait s comme des d chets m nagers Rapportez les au point de collecte appropri pour le recyclage 5 GARANTIE Notre garanti...

Страница 4: ...ves in particular LVD and EMC The rubbish bin with a line through it indicates that in the European Union the product must undergo selective disposal in compliance with Directive WEEE 2002 96 EC This...

Страница 5: ...or category exceeds those stated Do not use the instrument if it seems to be damaged incomplete or poorly closed Do not use the instrument in an explosive atmosphere or in the presence of flammable g...

Страница 6: ...diode and capacitance 21 2 4 Current 21 2 5 Non contact voltage detection NCV 22 2 6 Automatic standby 22 3 CHARACTERISTICS 23 3 1 Reference conditions 23 3 2 Electrical characteristics 23 3 3 Environ...

Страница 7: ...probes 2mm in diameter and 15mm long one red and one black 300V CAT II Figure 4 Test probes 4mm in diameter and 19mm long one red and one black 300V CAT II Figure 5 MiniFlex MA101 250 flexible current...

Страница 8: ...ds Five position rotary switch COM terminal Lead terminated by a probe tip holder 600 V CAT III 1000 V max COM DIGITAL TESTER Select C A 757 600 V CAT III 1 5 BACK When the instrument is not being use...

Страница 9: ...e instrument indicates a resistance near zero and emits a continuous beep Set the switch to V and measure a known voltage If the results of these four tests are correct your instrument is ready for us...

Страница 10: ...mode Above 3 M the display unit indicates OL The 30 M range is not available in automatic mode resistance mode must be used Diode In diode test place the red probe tip on the anode of the diode to be...

Страница 11: ...eps once a second and the backlighting flashes at the same rate the buzzer beeps twice a second and the backlighting flashes at the same rate the buzzer beeps three times a second and the backlighting...

Страница 12: ...dc to 2 999 Vdc 3 00 V to 29 99 V 30 0 V to299 9 V 300 V to 1000 V 100 mVac to 2 999 Vac Resolution 1 mV 10 mV 100 mV 1 V Intrinsic uncertainty in Vdc 2 3 pt Intrinsic uncertainty in Vac 3 4 pt Input...

Страница 13: ...ainty 5 10 pt 5 5 pt In these ranges subtract the no load value typically 50pF from all readings Range 30 F 300 F 3 mF 30 mF Measurement range 3 00 to 29 99 F 30 00 to 299 9 F 0 300 to 2 999 mF 3 00 t...

Страница 14: ...5 CHARACTERISTICSOFCONSTRUCTION C A 757 Dimensions L x l x P 180 x 52 x 45 mm Mass 200 g approx Cable length 142 cm Protection rating IP 54 according to IEC 60529 IK 04 according to IEC 50102 Drop tes...

Страница 15: ...e replacement of the batteries Spent batteries must not be treated as ordinary house hold waste Take them to the appropriate recycling collection point 5 WARRANTY Except as otherwise stated our warran...

Отзывы: