GB - User’s manual

C.A 6532
C.A 6534

Megohmmeters

Содержание C.A 6532

Страница 1: ...GB User s manual C A 6532 C A 6534 Megohmmeters...

Страница 2: ...optimize the effects of the product on the environment In particular this appliance exceeds regulation requirements with respect to recycling and reuse The CE marking indicates conformity with Europea...

Страница 3: ...ned Observe the environmental conditions of use Except for voltage measurements make no measurements on live devices Do not use the instrument if it seems to be damaged incomplete or poorly closed Bef...

Страница 4: ...2 8 HOLD function 19 2 9 Backlighting 20 2 10 SET UP 20 2 11 Alarm function 21 2 12 Automatic stop 22 2 13 Storage 22 2 14 Bluetooth communication 24 2 15 Errors 26 2 16 Resetting the instrument 27 3...

Страница 5: ...leads or accessories is the responsibility of the assembler of the system Each time before use always check that the insulation of the leads or accessories is in good condition Any element whose insu...

Страница 6: ...Bluetooth adapter DataView software 1 3 REPLACEMENT PARTS 2 straight right angle safety leads red and black 1 50 m long 2 crocodile clips red and black 2 test probes red and black 2 wire grips red an...

Страница 7: ...M MEM DARPI T1T2 k nF km M mA A G VHz G M HOLD REL P G nF F Connection terminals Backlit digital display unit 6 function keys Eight position switch to choose the function or to switch the instrument o...

Страница 8: ...50V 500V 25V TEST REL G M VDC AC k ALARM MEM k M mA A G VHz G M HOLD REL P Connection terminals Backlit digital display unit 6 function keys TEST button to start the measurements Ten position switch t...

Страница 9: ...9 1 4 3 ON THE BACK Captive quarter turn screw Battery compartment cover Prop Magnet for attachment to a metallic surface Non skid pads...

Страница 10: ...n the electronics industry It can also be used with suitable probes to test the immunity of walls and floors to electrostatic discharges ESD C A 6532 C A 6534 Test voltages for insulation measurements...

Страница 11: ...he durations of insulation measurements to choose the continuity test current and to program the alarm thresholds Rel The Rel key is used to display the measurement from which a stored reference measu...

Страница 12: ...nt 2 2 VOLTAGE MEASUREMENT Set the switch to V or to one of the M positions Start by making sure of the proper operation of the voltage measurement by measuring a known voltage before each use For exa...

Страница 13: ...est voltage and discharges the device being tested The symbol is displayed until the voltage on the device has fallen below 70 V Do not disconnect the leads and do not start any measurement while the...

Страница 14: ...the value displayed When the time is displayed press the X key to enter programming mode When the first digit blinks you can change it using the S key Press X to go to the next digit and S to change i...

Страница 15: ...ONTROL PROBE OPTION The remote control probe is used to trigger the measurement using the remoted TEST button on the probe To use this accessory refer to its operating instructions Interpretation of t...

Страница 16: ...y changes to zero and the symbol is displayed The resistance of the leads will be systematically subtracted from all continuity measurements If the resistance of the leads is 10 there is no compensati...

Страница 17: ...ts value R1 Then reverse the leads and note the value R2 Calculate the mean R R1 R2 2 M mA If an external voltage 15 V appears during the continuity measurement the instrument is protected without a f...

Страница 18: ...k MR OFF 50V M As for a continuity measurement connect the device to be tested to the terminals of the instrument The device to be tested must not be live see 2 4 3 G nF k nF km The instrument displa...

Страница 19: ...dition the measured value as a of the stored value M REL G In insulation measurements only the digital display is modified by the REL The bargraph continues to display the true measured value To exit...

Страница 20: ...t to OFF then X to validate the change The symbol disappears from the display when Set up is exited 2nd press on S P Automatic switching off is activated To deactivate it press X to make OFF blink S t...

Страница 21: ...displayed The symbol starts blinking you can change it to using the S key This symbol indicates the direction of the alarm threshold for a low threshold and for a high threshold Press the X key again...

Страница 22: ...made This automatic switching off can be disabled see 2 10 2 13 STORAGE 2 13 1 RECORDING A MEASUREMENT To record a measurement it is first necessary to freeze the display using the HOLD key or to wai...

Страница 23: ...mber has been chosen you can see all information concerning the measurement Long press thee MEM key then use the S key to scroll the information To exit from this rereading of records long press MEM a...

Страница 24: ...h to MR Long press the CLR key Press the S key and the record number is replaced by ALL To cancel long press the CLR key again Otherwise to confirm the erasure of all records long press the MEM key Th...

Страница 25: ...l is displayed and the instrument waits for a message from the computer When the link is set up the symbol starts blinking If your PC does not have a Bluetooth port install a USB Bluetooth adapter The...

Страница 26: ...again 2 15 1 PRESENCE OF A VOLTAGE BEFORE AN INSULATION MEASUREMENT G M mA A G If during an insulation measurement the value to be measured exceeds the measurement range which depends on the instrumen...

Страница 27: ...15 3 PRESENCE OF A VOLTAGE DURING A CONTINUITY RESISTANCE OR CAPACITANCE MEASUREMENT C A 6532 2 15 4 MEMORY FULL When the memory is full it is no longer possible to record measurements and the instrum...

Страница 28: ...reading R and in number of display points ct a R b ct 3 2 ELECTRICAL CHARACTERISTICS 3 2 1 VOLTAGE MEASUREMENTS Particular reference conditions Peak factor 1 414 in AC sinusoidal signal Specified mea...

Страница 29: ...ent range 0 1 M 50 G Resolution 9 segments per decade Intrinsic uncertainty 5 1 segment When the measurement range is exceeded the whole bargraph is displayed Test voltage With a test current 1 mA the...

Страница 30: ...200mA 0mA 20mA 20 mA range 20mA 5mA Specified measurement range 0 250 mA Resolution 1 mA Intrinsic uncertainty 2 2 ct Compensation of the leads 0 to 9 99 3 2 5 RESISTANCE MEASUREMENTS Specified measu...

Страница 31: ...humidity 20 to 80 RH V F 1 2 ct Frequency 15 3 to 800Hz V 1 2 1 ct Supply voltage 6 6 to 9 6V V F 0 1 2 ct Common mode rejection in AC 50 60 Hz 0 to 600VAC V 50dB 40dB 3 3 2 INSULATION MEASUREMENT Qu...

Страница 32: ...and 250V ranges from 10k to 3 G 6 2 ct 10 2 ct 500V range from 100k to 10G 6 2 ct 10 2 ct 0 to 1 F 50V range d 5G 250V range d 15G 6 2 ct 10 2 ct Common mode rejection in AC 50 60 Hz 0 to 600VAC V 50d...

Страница 33: ...indicated by the manufacturer E3 influence of the temperature between 0 and 35 C In continuity measurement B A 1 15 E1 E2 E3 3 5 POWER SUPPLY The instrument is powered by six 1 5 V alkaline AA LR6 ba...

Страница 34: ...AL STANDARDS The device is compliant per IEC 61010 1 and IEC 61010 2 030 600V CAT IV The device is compliant per IEC 61557 parts 1 2 4 and 10 3 9 ELECTROMAGNETIC COMPATIBILITY CEM The instrument is co...

Страница 35: ...e the instrument again until it is completely dry 4 2 REPLACING THE BATTERIES When the symbol starts blinking on the display unit the batteries must all be replaced Disconnect the unit completely and...

Страница 36: ...he following cases Inappropriate use of the equipment or use with incompatible equipment Modifications made to the equipment without the explicit permission of the manufacturer s technical staff Work...

Страница 37: ...37...

Страница 38: ...Fax 33 1 46 27 73 89 info chauvin arnoux com www chauvin arnoux com INTERNATIONAL Chauvin Arnoux Group T l 33 1 44 85 44 38 Fax 33 1 46 27 95 69 Our international contacts www chauvin arnoux com conta...