42
SPECIFICATIONS
GENERAL SPECIFICATIONS
Conformity to the EN 6110-1 Ed 95 standard
:
Instrument completely protected by double insulation
EN 61010-2-032 Ed 95, 150 V, category
III
and degree of pollution 2 (class 2).
Leakage line and distance in the air: 3.3 mm minimum
Emission
: NF EN 61326-1, ed 98
Immunity
: NF EN 61326-1, ed 98
Limiting overloads
:
permanent current 100 A max. (50/60 Hz)
transient currents (< 5 s) 200 A (50/60 Hz)
Case
: polycarbonate
Dimensions
: 55 x 100 x 240 mm
Max. Clamping diameter
: 32 mm Ø
Weight
: approx. 1 kg
Watertightness
: IP30, Group III equipment as per EN 60529 Ed 92
IK04, as per EN 50102 Ed 95
Power supply
: - 9 V alkaline battery 6LF22 or equivalent
- average consumption: approx. 40 mA
- average service life: approx. 8 hours,
i.e. 1000 measurements of 30 seconds.
With a rechageable NiCad battery, the average service life is
approx. 400 measurements of 30 seconds.
Note
:
extreme environmental conditions may disturb the internal microprocessor.
Simply disconnecting the battery may be sufficient to eradicate this malfunction.