8.7. My Bluetooth measurements are not displayed or are not refreshed in the
measurements manager
Make sure that acquisition has not been stopped. To preserve consistency between the image displayed and the
Bluetooth measurements read on the peripherals at the time of acquisition, the measurements are refreshed only
when there is a new acquisition. If a peripheral is connected, but acquisition has been stopped, its measurements
will not be displayed until there has been another acquisition.
Following an incorrect reading, the measurement read and the measurement expected may not be consistent. This
can prevent refreshing. To eliminate the ambiguity, change the measurement function on the peripheral (rotary
switch on clamps) and return to the initial position.
If the problem persists:
1. Switch off the Bluetooth peripheral.
2. Wait until this disconnection is detected on the camera.
3. Switch your peripheral back on (Bluetooth mode activated).
4. Press
F2
to try again to connect the instrument from the Bluetooth peripherals manager
(Main/Parameters/Bluetooth/Peripherals menu).
If continuous acquisition is activated, your measurements must appear in the Bluetooth measurements
manager (Main/Parameters/Bluetooth/Measurements menu).