  The  backlight  turns  ON  automatically  when  entering  a  function  and 

turns OFF after the function is complete. Press the

 DISPLAY

 button to turn it back 

on.

Figure 2-2

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание AEMC Instruments 2135.01

Страница 1: ...DIGITAL GROUND RESISTANCE AND SOIL RESISTIVITY TESTER E N G L I S H User Manual www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 3: ...during part of this recharging operation Only use the charger supplied with your instrument Use of another charger may be dangerous Only charge your instrument at temperatures between 0 and 40 C manu...

Страница 4: ...ce Conditions 15 3 2 Electrical 15 3 2 1 Frequency Measurements 15 3 2 2 Voltage Measurements 15 3 2 3 DC Resistance Measurements 16 3 2 4 AC Earth Ground Resistance Measurements 17 18 3 3 Data Storag...

Страница 5: ...ing the Distance for Electrode Placement 35 36 5 6 5 Schlumberger Method 37 6 1 Switching the Test Voltage 39 6 2 Measurement Frequency Selection 40 6 3 Manual Settings for m Measurements 40 6 3 1 Con...

Страница 6: ...e Report to a PDF File 54 9 1 Maintenance 55 9 2 Cleaning 55 9 3 Fuse Replacement 55 9 4 Recharging the Battery 56 9 5 Replacing the Battery 57 10 1 Error Reporting 50 10 1 1 Electrode Resistance Too...

Страница 7: ...eption surement category Do not use the instrument or its accessories if they appear damaged greater than or equal to those of the instrument CAT IV 50V Use only Check that no terminal is connected an...

Страница 8: ...ag 110 240V power adapter with US power cord optical USB cable rechargeable NiMH battery pack and a USB stick with DataView software ground tester workbook and user manual Ground Resistance Tester Mod...

Страница 9: ...oded leads on spools red blue one 30 ft lead green two 5 ft color coded leads red blue two T shaped auxiliary ground electrodes set Test Kit for 4 Pole 4 Point Testing 300 ft Cat 2135 36 Includes one...

Страница 10: ...more than 250VAC 250VAC against dust and water when closed the test button is also sealed against the elements manual and automatic test frequency selection from 40 to 513Hz user optional 12V battery...

Страница 11: ...ptimum test ac Stores up to 512 complete test results in internal memory Optically isolated USB communication Remote set up and operation of all measurements using DataView software supplied Automatic...

Страница 12: ...Figure 2 1 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 13: ...automatically when entering a function and turns OFF after the function is complete Press the DISPLAY button to turn it back on Figure 2 2 www GlobalTestSupply com Find Quality Products Online at sale...

Страница 14: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 15: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 16: ...a sampling frequency of 4028Hz low pass FFT The frequency of the strongest spectral component is displayed Display updates approx 3 times per s between 70 and 75V at terminals H Z and E X If terminal...

Страница 17: ...urement Standard EN 61557 part 4 16VDC if resistance 22 reduced to 10VDC 200mADC for resistances 20 50Vrms protection up to 250V is guaranteed 2 H 60Vpeak 10Hz approx 5 s 3 per s in manual mode Possib...

Страница 18: ...operating error of AC resistance measurements can be less than that speci are matched to those of the current channel Measurement of auxiliary electrodes RH RS RES RE The START button must be pressed...

Страница 19: ...Pole earth resistance measurement RE OUT H Voltage Current measurement EN 61557 part 5 Selectable from 41 to 128Hz see table in 6 2 200mAAC test frequency 250Vrms www GlobalTestSupply com Find Quality...

Страница 20: ...ce of the earth rods R R R R assumed to be identi cal OUT rod Measurement of Auxiliary Electrodes R H R S R ES R E The START button must be pressed for more than 2s For frequencies between 41 and 128H...

Страница 21: ...y isolated USB Rechargeable 9 6V 3 5Ah NiMH Battery Pack 110 220 50 60Hz external charger with 18VDC 1 9A output or 12VDC 0 63A 250V 5x20mm 1 5kA www GlobalTestSupply com Find Quality Products Online...

Страница 22: ...4mm recessed banana jacks 32 to 113 F 0 to 45 C 0 to 90 RH 1 0 to 95 F 0 to 35 C 0 to 75 RH 3000m Electromagnetic Compatibility Measurement according to EN 61557 CAT IV 50V Pollution Degree 2 www Glob...

Страница 23: ...rument takes the measurement and stops automatically using the DISPLAY button MEM button see 7 1 Set the switch to the desired function Make the connections appropriate to the function Select MANUAL m...

Страница 24: ...see 7 SET UP PUSH button message appears on the display Various parameters are accessible by pressing the corresponding button using the SELECT button and CHANGE buttons To exit SET UP turn the rotar...

Страница 25: ...ress the Hz OPTIONS age Uout blink Hz OPTIONS again 1 Set the switch to 2 Connect the resistance to terminals H Z and E X www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSuppl...

Страница 26: ...sation subtracts the resistance of the test leads from the measured result Short the two measurement leads connected to H Z and E X terminals Press the 2nd button then the START STOP button to start t...

Страница 27: ...e changed using the SET UP Turn the alarm ON or OFF Access the type of alarm Choose a low or high alarm between 1 and 999 Complete the adjustment of the alarm setting and www GlobalTestSupply com Find...

Страница 28: ...ss Hz OPTIONS once again H and E will blink H S ES and E will all blink Hz OPTIONS again 1 Set the switch to 2 Connect the resistance to all 4 terminals www GlobalTestSupply com Find Quality Products...

Страница 29: ...electrodes H Z and S Y to form a straight line with the earth electrode under test The distance between electrode S Y and the earth electrode is 62 of the distance between electrode H Z and the earth...

Страница 30: ...The hundredths will blink 4 5 6 7 To terminate the programming of distance press the 2nd button and then the DISTANCE button It is also necessary to program distance A This is done in the same way as...

Страница 31: ...a long press of the START STOP button RH RS will be displayed 2s Repeat the process at 52 and at 72 for the S electrode ing the S Y H Z and E X and plotting the results A distinct plateau should resu...

Страница 32: ...few percent If this If not electrode S Y necessary to increase distance d for the H Z auxiliary electrode and repeat the measurements Positioning of the Auxiliary Electrodes To make sure that your ea...

Страница 33: ...reel to keep the cables as far apart as possible on the ground metallic conductors cables rails fences etc Connect the cables to terminals H Z and S Y disconnect the earth strap from the rod under te...

Страница 34: ...measure the resistances of auxiliary electrodes H Z and S Y or if the resis tance of the electrodes is too large see 10 1 start the measurement with a long press of the START STOP button RH RS U will...

Страница 35: ...soil in question which can be useful for geological analysis the exploration of deposits hydrological studies etc and to determine the location of an earth electrode www GlobalTestSupply com Find Qua...

Страница 36: ...press the Hz OPTIONS button twice x 2 The distance can be programmed before or after the measurement If it is not remains indeterminate 1 Set the switch to 2 Press the 2nd button then the DISTANCE but...

Страница 37: ...n a straight line at a distance d from one an other and at a depth p 1 20 d Distance d must be between 2 and 30m 3 Connect the cables to the electrodes then to terminals H Z S Y ES Xv and E X in seque...

Страница 38: ...START STOP button R and R then R and R will be displayed 2s 1 Set the switch to 2 Place electrodes S Y and ES Xv at a distance A apart then place elec trodes H Z and E X in the same straight line at...

Страница 39: ...rement parameters press DISPLAY S R d U I and its frequency U and its frequency To measure the resistances of the auxiliary electrodes H Z S Y ES Xv and E X or if the resistance of the electrodes is t...

Страница 40: ...TO indicator blinks AUTO to MANUAL MANUAL mode you can press Hz OPTIONS repeat ment function selected START STOP button the circular arrows on the display will rotate the measurement can be stopped by...

Страница 41: ...sing the Hz OPTIONS button until the frequency flashes USr 55 92 110 119 128 Hz 128 Hz is default and Pressing Hz OPTIONS in MANUAL mode allows the following parameters to be Terminal symbols H and E...

Страница 42: ...function must be selected The alarm threshold must be low Pressing on Hz OPTIONS in MANUAL mode allows the following parameters to EARTH blinks Earth Coupling Measurement 128 Hz blinks Test Frequency...

Страница 43: ...surement of R2 For this second H Z and S Y electrodes in the same positions as Store this result in the same memory location as before by pressing the MEM button twice The screen now displays the mess...

Страница 44: ...een 16 and 32V Pressing the Hz OPTIONS button allows the following parameters to be changed Switch to S Schlumberger method 128Hz blinks Change the test frequency Switch between 16 and 32V ment result...

Страница 45: ...additional addressing criterion so each occupies only one memory address before any storage in memory see 4 2 Press the MEM button The tester will automatically suggest the next free memory location...

Страница 46: ...s chosen View software see 8 Push once on the MEM records Push a second time on the MEM button The display will show dEL ALL Change the blinking NO to YES button then perform a long press 2s on the ME...

Страница 47: ...nd hold down 2s the MEM button To exit without erasing perform a short press on the MEM button Note that if the test stored in this Object Test number is a sweep test which con deletes a single measur...

Страница 48: ...pears on your screen labeled DataView 2 3 The DataView setup screen appears Setup interface All Setup screens and diaogs will immediately appear in the selected language DataView software you can sele...

Страница 49: ...eports Core click Next 6 The Setup program now informs you that it is ready to install DataView If you return to earlier screens Otherwise click Install to begin installation 7 of the software is alre...

Страница 50: ...www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 51: ...of the connection Date and time from the clock Battery status cation This window defaults to the instrument mode that is selected If the indicated items are not shown on the screen select Restore Defa...

Страница 52: ...status at the input of the meter is always displayed 3 Pole Ground Resistance AC Distance d from terminal X to terminal Y Distance a from terminal X to terminal Z table selection www GlobalTestSupply...

Страница 53: ...AC table selection 4 Pole Test Resistivity AC lumberger test method between the test electrodes needs to be entered sweep table selection www GlobalTestSupply com Find Quality Products Online at sale...

Страница 54: ...is stable Others generally when there is a manual fre quency show a STOP Click the plus sign next to Recorded Data from the instrument in the Instrument Prints the data Opens the newly created report...

Страница 55: ...s The Operator and Site tabs allow you to maintain lists of operators and sites in displaying a report Only a single set of Custom parameters can be maintained unlike the Operator and Site lists Recor...

Страница 56: ...ect the unit from all circuits before opening the case impossible to perform resistance measurements To test this fuse turn the switch to the m leads to terminals H Z and E X and start a measurement I...

Страница 57: ...attery charger 5 6 Put the unit back into the housing and screw it into place Use only the battery charger supplied with the unit other chargers may be hazardous Recharge the battery only at temperatu...

Страница 58: ...eft connector AC power source suitable for Europe and the U S The frame of the battery symbol will DC car outlet with a special charging unit The battery in this unit is a special product with customi...

Страница 59: ...inal for the optical circuitry or components 10 Return the back of the unit with the new battery in the case to the front panel with the components taking care not to stretch the battery leads Careful...

Страница 60: ...press on the START STOP but them to display the correct result or out of range and the smoothing function SMOOTH may correct this Press the 2nd button then the SMOOTH smoothing needs to be either conn...

Страница 61: ...the terminals and that the measurement is impossible active measurement means that the operat there is an incorrect connection cates that measurement current IES or ISEL is too low interfere with the...

Страница 62: ...ir will be necessary if the error persists Error 18 indicates that the rechargeable battery in the tester cannot be recharged If this error occurs during recharging disconnect the tester from the char...

Страница 63: ...th potential ence earth electrode Earth electrode ES Xv A connection on the earth electrode or earthing system used for measuring the potential of the earth electrode Earth resistance The resistance m...

Страница 64: ...de An additional earth electrode earth stake earth rod Touch voltage the resulting current being limited by the body and the local earth resistance earth electrodes earth connections or entire earthin...

Страница 65: ...terminal H to E R Act 1 R1 sockets R1 RA C R2 earth sockets R2 Rb C R1 2 earth sockets R RA b RA A R1 C Rb b R1 C RC A and Rb Rc R1 2 2 RE RH RPASS measurement with interference current circulating i...

Страница 66: ...E US ES USR USrEL S method expressed as m W expressed as m Act becomes In 2 In this case the resistances of the 4 rods used for measurement are indicated by R R R R www GlobalTestSupply com Find Qual...

Страница 67: ...ation traceable to N I S T Inc d b a AEMC Instruments Or contact your authorized distributor Costs for repair standard calibration and calibration traceable to N I S T are If you are experiencing any...

Страница 68: ...our records What AEMC If a malfunction occurs within the warranty period you may return the instrument or a proof of purchase AEMC Instruments will at its option repair or replace the faulty material...

Страница 69: ...Chauvin Arnoux Inc d b a AEMC Instruments www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...