Neo Stimulation Channel 1 & 2 + EMG Module  Manual

•  This unit should be operated at 10°C to 45°C and 0% to 90% Relative Humidity. 

The unit should be transported and stored at 0°C to 60°C and 0% to 95% Relative 
Humidity.

•   Device is designed to comply with electromagnetic safety standards.  This 

equipment generates, uses, and can radiate radio frequency energy and, if not 
installed and used in accordance with instructions, may cause harmful 
interference to other devices in the vicinity.  However, there is no guarantee that 
interference will not occur in a particular installation. Harmful interference to 
other devices can be determined by turning this equipment on and off. Try to 
correct the interference using one or more of the following:

-  Reorient or relocate the receiving device
-  Increase the separation between the equipment
-  Connect the equipment to an outlet on a different circuit from that to 

which the other device(s) are connected and consult the factory field 
service technician for help.

-  Consult your authorized DJO dealer for help. 

•   Do not operate this unit when connected to any unit other than DJO devices or 

accessories specifically described in user or service manuals.

•   DO NOT disassemble, modify, or remodel the unit or accessories. This may cause 

unit damage, malfunction, electrical shock, fire, or personal injury.

•   DO NOT permit foreign materials, liquids or cleaning agents to enter the unit, 

including, but not limited to, inflammables, water, and metallic objects from 
entering the unit, to prevent unit damage, malfunction, electrical shock, fire, or 
personal injury.

•   If you have difficulty operating the unit after carefully reviewing this user 

manual, contact your DJO dealer for assistance.

DANGER

CAUTION

WARNING

•  Stimulus delivered by the TENS waveforms of this device, in certain 

configurations, will deliver a charge of 25 microcoulombs (μC) or 
greater per pulse and may be sufficient to cause electrocution. 
Electrical current of this magnitude must not flow through the 
thorax because it may cause a cardiac arrhythmia.

•  Be sure to read all instructions for operation before treating patient.
•  Make certain the unit is electrically grounded by connecting only to 

a grounded electrical service receptacle conforming to the applicable 
national and local electrical codes.

•  Care must be taken when operating this equipment around other 

equipment. Potential electromagnetic or other interference could occur 
to this or to the other equipment. Try to minimize this interference by not 
using other equipment in conjunction with it.

•  The safety of TENS waveforms for use during pregnancy or birth has not 

been established.

•  TENS is not effective for pain of central origin. (This includes headache.)
•  TENS waveforms have no curative value.
•  Electronic monitoring equipment (such as ECG monitors and ECG alarms) 

may not operate properly when electrical stimulation is in use.

•  TENS is a symptomatic treatment, and as such, suppresses the sensation 

of pain which would otherwise serve as a protective mechanism.

•  Contaminated electrodes and leadwires can lead to infection.
•  Use of electrode with degraded hydro-gel can result in infection to the 

skin burn.

•  Use of electrode on multiple patients can lead to infection.
•  Stop treatment immediately if patient experiences discomfort or pain.
•  Powered muscle stimulators should be used only with the leads and 

electrodes recommended for use by the manufacturer.

•  In the event of all 300-Level or a 200-Level error message that cannot be 

resolved, immediately stop all use of the system, and contact the dealer 
or DJO for service. Errors and Warnings in these categories indicate an 
internal problem with the system that must be tested by DJO or a Trained 
Technician before any further operation or use of the system.

-  Use of a system that indicates an Error or Warning in these categories 

may pose a risk of injury to the patient, user, or extensive internal 
damage to the system.

•  DO NOT connect the unit to an electrical supply without first verifying that 

the power supply is the correct voltage. Incorrect voltage may cause unit 
damage, malfunction, electrical shock, fire, or personal injury. Your unit 
was constructed to operate only on the electrical voltage specified on the 
Voltage Rating and Serial Number Plate.

-  Contact your DJO dealer if the unit is not properly rated.

•  Keep electrodes separated during treatment. Electrodes in contact with 

each other could result in improper stimulation or skin burns.

•  Long term effects of chronic electrical stimulation are unknown.
•  Stimulation should not be applied over the anterior neck or mouth. 

Severe spasm of the laryngeal and pharyngeal muscles may occur and the 
contractions may be strong enough to close the airway or cause difficulty 
in breathing.

•  Stimulation should not be applied transthoracically in that the 

introduction of electrical current into the heart may cause cardiac 
arrhythmia.

•  Stimulation should not be applied over swollen, infected, and inflamed 

areas or skin eruptions, e.g., phlebitis, thrombophlebitis, varicose veins, 
etc.

•  Stimulation should not be applied over, or in proximity to, cancerous 

lesions.

•  Electrotherapy output current density is related to electrode size. Improper 

application may result in patient injury. If any question arises as to the 
proper electrode size, consult a licensed practitioner prior to therapy 
session.

•  Output current density is related to electrode size. Improper application 

may result in patient injury. If any question arises as to the proper 
electrode size, consult a licensed practitioner prior to therapy session.

Содержание Neo Stimulation Channel 1 & 2

Страница 1: ...el 1 2 EMG Module 13 8905 Rev F For complete User Operating Instructions including Cautions Warnings Dangers Indications and Contraindications refer to the User s Manuals Domestic Market US only refer toVectra Neo ClinicalTherapy System User Manual Item 13 7646 CDVersion item 13 7647 International Market refer to Intelect Neo ClinicalTherapy System User Manual Item 13 7651 CDVersion item 13 7652 T...

Страница 2: ... required not included 2 Phillips screwdriver Standard slotted screwdriver The system is programmed to automatically recognize the new Module therefore no software installation is required Complete the following steps to install the module 1 Power off device Remove main power cord from the rear of the device 2 Remove the blank faceplate over the second slot from the front on the left and right sid...

Страница 3: ...fer to Intelect Neo Clinical Therapy System User Manual Item 13 7651 CD Version item 13 7652 To obtain a copy contact your local representative or DJO Customer Care 6 Using a 2 Phillips screwdriver secure the module with the provided 8 32 x 5 16 screw at the bottom tab as shown 7 Insert the faceplate with connection openings supplied with module on the Left side at the bottom and snap into place a...

Страница 4: ... ECG monitors and ECG alarms may not operate properly when electrical stimulation is in use TENS is a symptomatic treatment and as such suppresses the sensation of pain which would otherwise serve as a protective mechanism Contaminated electrodes and leadwires can lead to infection Use of electrode with degraded hydro gel can result in infection to the skin burn Use of electrode on multiple patien...

Отзывы: