14
INTELECT
®
RPW 2 QUICK START GUIDE
• This device should be operated at 10°C to +40°C and 30% to 75% Relative Humidity (no condensation).
The device should be transported and stored at -20°C to +60°C and 10% to 90% Relative Humidity (no
condensation).
• Atmospheric pressure requirements: 70 kPa to 106 kPa.
• The Falcon handpiece should be operated at 10°C to +40°C and 5% to 95% Relative Humidity (no
condensation). The handpiece should be transported and stored at 0°C to +60°C and 5% to 95% Relative
Humidity (no condensation).
• The V-Actor HF handpiece should be operated at 10°C to +30°C and 5% to 95% Relative Humidity (no
condensation). The handpiece should be transported and stored at 0°C to +60°C and 5% to 55% Relative
Humidity (no condensation).
• The time required for Intelect RPW 2 to warm from the minimum storage temperature between uses until
Intelect RPW 2 is ready for its intended use when the ambient temperature is 20°C is 5 hours.
• The time required for Intelect RPW 2 to cool from the maximum storage temperature between uses until
Intelect RPW 2 is ready for its Intended use when the ambient temperature is 20 °C is 5 hours
• Under single fault condition and extreme usage conditions, the maximum temperature of the hand piece
can reach 47°C. There is no particular health risk associated with this temperature besides your comfort.
• Use of parts or materials other than DJO’s can degrade minimum safety.
• Connect to this device only items and equipment that have been specified in this IFU as part of the Intelect
RPW 2 System or that have been specified as being compatible with the Intelect RPW 2 System.
• DO NOT disassemble, modify, or remodel the device or accessories. This may cause device damage,
malfunction, electrical shock, fire, or personal injury.
• Before using the therapy device, the operator must ascertain that it is in correct working order and
operating condition. The cables and connectors, in particular, must be checked for signs of damage. The
device cannot be used if any part is damaged as there is a risk of electrical shock.
• Before each use always offer the patient hearing protection.
• Apply a sufficient amount of coupling gel to the patient’s skin in the treatment area and to the shock
transmitter.
• The maximum energy level used during treatment must not cause the patient undue pain under any
circumstances.
• DO NOT apply more than 300 shocks to the same spot during treatment.
• The surface of the transmitter can reach 47°C. Extended skin contact can cause minor burns. Interrupt
therapy after a maximum of 6000 pulses to allow the transmitter to cool.
• Operating the device at pressures higher than 3 bar without an impact surface can result in damage to the
handset.
• To avoid a trip hazard, keep all accessories and their cords separated during treatment by using the hook
located on the side of the device.
• Disconnect the instrument from the mains before starting any cleaning or overhaul work.