Drug Information
– Consult directions for the use of the drug/compound before delivery. Some drugs/compounds require a
specific polarization for use. Observe the indications, contraindications, warnings and precautions related to this issue.
•
Single Use
– Do not use electrodes that have been previously used. These patches have been designed for single use only.
Reuse may cause burns to patient.
•
Packaging Information
– Inspect the electrode packaging before use. Do not use electrodes past the Best If Used By date
printed on the packaging. Do not use electrodes if the packaging is not sealed. The electrodes may not be as effective if they
are used past the Best If Used By Date or if they have been exposed to high humidity for an extended period of time.
•
Electrode Inspection
– Inspect the electrodes before use. Discard any electrode that shows signs of alteration or damage, as
these electrodes may not be safe for use.
•
Excessive Motion
– The electrodes can be worn during normal activity. However, excessive motion where the electrodes have
been placed can cause poor contact between the skin and the electrode or uneven distribution of current, resulting in greater
risk of skin irritation.
•
Erythema
– A transient erythematous reaction, characterized by a uniform red pattern, can sometimes occur directly under
the electrodes. The redness usually disappears within a few hours after treatment. Advise the patient of this possibility before
starting treatment.
•
Positive Polarity with Chloride Ions
– Use only saline ampoule supplied for the return pad. For positive polarity (+), use only
drugs with Chloride (Cl-) counter ions. Use of tap water or any other solution may cause tattooing or staining.
•
Handle with Care
– Handle the system with care. Do not immerse the system in fluids or allow it to be connected with other
electrical devices. Do not drop, abuse, or in any way exceed normal use. Do not sterilize. Misuse may cause the device to
malfunction in ways that may lead to shock or burns.
•
Use with RF Equipment
– Do not operate this system in an environment where other devices are being used that intentionally
radiate electromagnetic energy in an unshielded manner. Portable and mobile RF communications equipment can affect
Medical Electrical Equipment.