Maintaining the Vectra Genisys Laser Module
The following items should be checked at least monthly to ensure proper operation of this unit:
• Power cord and plug: Check to make sure the cord is not frayed, kinked, and does not have torn or cut insulation.
• Applicator cable: Check to make sure the cable is flexible, free of kinks, not frayed, and the insulation is intact.
• Applicator lens: Check to see that there is no build-up of oil or foreign material on or behind the applicator lens.
Cleaning
To clean the accessories, use only soap and water.
The Vectra Genisys Laser Module’s case may be cleaned by wiping with a damp cloth or mild cleaning solution. Avoid abrasive
cleansers.
Service
The Vectra Genisys Laser Applicator must be recalibrated annually. Chattanooga Group laser products must be returned to the
factory or an authorized servicing dealer for repairs or recalibration. Should the Vectra Genisys Laser unit require service, contact the
selling dealer or Chattanooga Group Service Department.
Before each use, clean the plastic lens with NOVUS® Polish System (www.novuspolish.com). Make certain to apply with a clean
cloth. Failure to clean the lens between patient therapy sessions could cause laser light beam fragmentation, which may reduce
the effectiveness of the treatment.
WARNING