MVE TS Quick Reference Guide
A & B and Inlet (Hot Gas Bypass)
The following section describes how to adjust temperature alarm settings. At any time
during the following procedure, the user may exit the menu by pressing the
“EXIT” button
to return to the
“monitor” display mode. After 60 seconds of inactivity, the controller will
automatically return to the “monitor” display mode.
NOTE: Security Level 2 or higher is required to adjust temperature settings (see
“Password and Security Setup” section for details).
To exit any menu screen and return to the previous menu, press “<” key.
1.
Press “Setup”
Controller will prompt for a password. Type in the password using the number pad that appears and
press “Enter”.
2.
Press “Temperature Settings”