• Periodic Audits of Installed Products - Surface resistance 

  exponent number illuminates immediately and displays 

  Temperature and Relative Humidity during 15 second   

  electrification period making numerous measurements and 

  calculations to the display mantissa i.e LED displays 8 and 

  LCD displaying 7.14 as mantissa = 7.14 x 10E8 ohms or 

  714,000,000 ohms

• Laboratory Test Evaluation of Product - Display same as 

  above Periodic Audit

The Surface Resistance Test Kit is NIST calibrated and is 

available in two models:

PPE-5075.E

 July 2012 Page 1 of 6

Surface Resistance Test Kit 
Operation and Maintenance

TECHNICAL BULLETIN PPE-5075.E

Made in the

United States of America

© 2012 CHARLESWATER LTD.

Unit 17. Millbrook Business Park, Sybron Way • Crowborough, East Sussex TN6 3JZ United Kingdom

Phone: 00 44 (0) 1892-665313, Fax: 00 44 (0) 1892-668838 • Internet: 

Charleswater.co.uk

Figure 1. Item 

99105

 Surface Resistance Kit

Description

The Surface Resistance Test Kit is a portable, accurate, 

and versatile instrument designed to measure point to point 

resistance (Rp), resistance to ground (Rg), and surface 

resistivity. It is designed to perform test methods described in 

EN 61340-5-1 Annex A.  Features include:
•  Resistance measuring accuracy ± 10% (E11 and greater 

  ± 20%)

•  Resistance range <1.0 x 10E3 ohms to >10E12 ohms

•  Open circuit voltages of 10 and 100 ± 5%

•  Electrification period of 15 seconds

•  Electrodes 2.27 kilograms ± .06 kg with 50-70 durometer 

conductive pads

In addition, the Meter measures ambient temperature and 

relative humidity.

The Surface Resistance Test Kit (or its Meter) is referenced 

and designed to be used to make measurements in accordance 

with the test methods in EN 61340-5-1 Annex A Resistance 

Measurement Test method:
•  A. 1 Floor, working surface or storage rack working surface

•  A.2 Seating

•  A.3 Garments

•  A.5.4 Gloves, finger cots and tools (Neither 20 mm plates nor 

clip connectors provided)

Versatility:

Measures Rp, Rg or Resistivity (requires optional Resistivity 

Attachment, item 

99107

)

• Quick Checks - Surface resistance exponent number 

  illuminates immediately i.e. 8 = 10E8 ohms or 100,000,000 

  ohms

Model  Description 

 

      Power

99105 

Surface Resistance Test Kit         

2 x AA Batteries

99106 

Meter and Attachment 

       

2 x AA Batteries

99107 

Resistivity Attachment 

       N/A

A pair of Test Leads (item 

99108

can be bought separately. 

The Charleswater 

99109

 replacement electrodes can be 

purchased separately also.

Inspection

Remove the meter from the carton and inspect for damage. 

Each Surface Resistance Test Kit should include the following:

1  Protective carrying case

1  Meter

2  Test leads

2  2.27 kilogram electrodes

2  AA alkaline batteries

1  Certificate of NIST calibration

Properly store the Meter and component assemblies when not 

in use. 

ESD protective products should be tested:

A.  Prior to installation to qualify for listing in user’s ESD control 

plan approved ESD protective materials (see EN 61340-5 

paragraph 9.3.2 Qualified protective items)

B.  During initial installation

C. For periodic audits of installed products as part of EN 

61340-5-1 paragraph 9 Quality responsibilities 

In addition: Per EN 61340-5-2 paragraph 9.4 “When a new EPA 

is established, or an existing EPA is reconfigured, it should be 

thoroughly checked by the ESD co-ordinator and a certificate … 

issued.”

Electrification Period

The Surface Resistance Test Kit provides the proper 

electrification period of 15 to 20 seconds per ESD EN 

61340-5-1, after numerous readings and calculations 

are executed, then displays surface resistance mantissa 

measurement (Note: most analog type meters display 

measurements instantaneously).

Содержание 99105

Страница 1: ...ance Measurement Test method A 1 Floor working surface or storage rack working surface A 2 Seating A 3 Garments A 5 4 Gloves finger cots and tools Neither 20 mm plates nor clip connectors provided Versatility Measures Rp Rg or Resistivity requires optional Resistivity Attachment item 99107 Quick Checks Surface resistance exponent number illuminates immediately i e 8 10E8 ohms or 100 000 000 ohms M...

Страница 2: ...ponent or power of the number is displayed immediately with LED illuminating number If 8 is illuminated the measurement is in the 10E8 or 10E8 ohm range To make a measurement the button is to be depressed for 15 20 seconds for the unit to make numerous readings and calculations Testing in accordance with EN 61340 5 1 requires 15 to 20 seconds of electrification in contrast most analog type meters ...

Страница 3: ...ectrode on different sections Clean surface for Laboratory test but do not initially clean surface for installed products if fails clean and retest Laboratory Test Procedure Guideline For laboratory test of ESD Working surfaces Floor Materials Footwear Garments or Seating best advice is to follow test methods of Annex A of EN 63140 5 1 The test specimens and electrodes should be cleaned twice with...

Страница 4: ...cally perform a lot sampling at incoming inspection Floor Per EN 61340 5 1 paragraph 10 1 These audits shall be performed at intervals not exceeding those defined by the ESD co ordinator Users typically perform these quarterly Seating Per EN 61340 5 1 paragraph 10 1 These audits shall be performed at intervals not exceeding those defined by the ESD co ordinator Users typically perform these quarte...

Страница 5: ...ate to 2 5 except at 10E11 and 10E12 5 Thermometer accurate to 1 F Humidity meter accurate to 2 Test leads 99 Isopropyl alcohol and cleaning wipes SET UP A Test Area Area needs to be free of any high voltage transformer or power supply Not under any type of fluorescent lighting or high power lighting B Worksurface needs to be covered with conductive mat at 1 0 x 10E3 or less connected to earth gro...

Страница 6: ...d or replaced at Charleswater s option free of charge Call Customer Service at 00 44 0 1892 665313 for a Return Material Authorisation RMA and for proper shipping instructions and address You should include a copy of your original packing slip invoice or other proof of purchase date Any unit under warranty should be shipped prepaid to the Charleswater factory Warranty replacements will take approx...

Отзывы: