B. EMC Guidance and Manufacturer's Declaration 

Guidance and manufacturer’s declaration-electromagnetic emissions

 

The device is intended for use in the electromagnetic environment specified below. The 
customer or the user of the device should assure that it is used in such an environment. 

Emission test 

 

Compliance 

 

Electromagnetic 
environment-guidance 

 

RF emissions CISPR 11   

Group 1   

The device uses RF energy only for 
its internal function. Therefore, its 
RF emissions are very low and are 
not likely to cause any interference 
in nearby electronic equipment.   

RF emissions CISPR 11   

Class A   

The product is suitable for use in all 
establishments other than domestic 
and those directly connected to a 
low voltage power supply network 
which supplies buildings used for 
domestic purposes. 

 

Guidance and manufacturer’s declaration-electromagnetic immunity

 

The device is intended for use in the electromagnetic environment specified below. The 
customer or the user of the device should assure that it is used in such an environment.   

Immunity test

 

IEC 60601 

test level

 

Compliance 

level

 

Electromagnetic 

environment-guidance

 

Electrostatic 

discharge(ESD)   

IEC 61000-4-2   

±

  8 kV contact 

±

  2 kV, ±  4 kV, 

±  8 kV, ±  15 kV 

air 

±

  8 kV contact 

±

  2 kV, ±  4 kV, 

±  8 kV, ±  15 kV 

air 

Floors should be wood, concrete 
or ceramic tile. If floors are 
covered with synthetic material, 
the relative humidity should be 
at least 30%   

Power 
frequency(50/60 
Hz)  magnetic field 
IEC 61000-4-8   

30 A/m   

30 A/m   

The device power frequency 
magnetic fields should be at 
levels characteristic of a typical 
location in a typical commercial 
or hospital environment. 

NOTE UT is the a.c. mains voltage prior to application of the test level. 
 

 

 

Содержание HM80D

Страница 1: ...1 USER MANUAL HM80D Portable Infant Height Stadiometer Please keep the instruction manual at hand all the time for future reference...

Страница 2: ...ructions for use Medical electrical device Type B applied part Device catalogue number model number Name and address of authorized representative in the European Union Device is a medical device Text...

Страница 3: ...national copyright law All content is licensed and usage is subject to written authorization from Charder Electronic Co Ltd hereinafter Charder Charder is not liable for any damage caused by a failure...

Страница 4: ...S21NEOV MS4200 Infant Scales 10 III Assembly with MS5900 Infant Scale 12 IV Assembly with MS2400 Infant Scale 15 V Assembly with MS5980 Infant Scale 17 VI Preparing for Use 18 Keys 18 Inserting Batter...

Страница 5: ...ed issues General Handling Device should be placed on stable flat solid non slippery surface Ensure all parts are properly locked and tightened before operating the device Measurement accuracy require...

Страница 6: ...chemical electrochemical or electrical interference unless damage is attributable to negligence on the part of Charder This device does not contain any user maintained parts All maintenance technical...

Страница 7: ...which supplies buildings used for domestic purposes Guidance and manufacturer s declaration electromagnetic immunity The device is intended for use in the electromagnetic environment specified below...

Страница 8: ...d 2 3 P 800MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m F...

Страница 9: ...ces are controlled The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transm...

Страница 10: ...3500 MS21NEOV MS4200 Infant Scales 1 Attach bracket to tray using bracket screws 2 2 Slide HM80D Height Stadiometer onto bracket until you hear a clicking noise that indicates it has been securely att...

Страница 11: ...11 3 Assembly complete To detach HM80D press on tab at the bottom while pulling...

Страница 12: ...12 III Assembly with MS5900 Infant Scale 1 Push bracket holder cover towards screen 2 Remove bracket cover...

Страница 13: ...13 3 Attach SS 6611 bracket to device with two screws 4 Slide HM80D Height Stadiometer onto bracket until you hear a clicking noise that indicates it has been securely attached...

Страница 14: ...14 5 Place bracket holder cover back into place To detach HM80D remove bracket holder cover and press on tab at the bottom while pulling...

Страница 15: ...ce and fasten screws using screwdriver Version with rubber head pad Version without rubber head pad 2 Connect height measure attachment HM80D or HM80M to bracket A clicking noise will be heard Attach...

Страница 16: ...16 3 To detach height measure find the latch located at the rear of the attachment Press down on the buckle and gently remove the attachment from bracket...

Страница 17: ...17 V Assembly with MS5980 Infant Scale 1 Attach L shaped height rod bracket to column with two screws 2 Complete installation of height stadiometer...

Страница 18: ...ate to lowest value UNIT Switch measurement unit between imperial and metric HOLD Hold height measurement value on screen ON OFF Turn device on and off Inserting Batteries Insert AAA batteries 2 Calib...

Страница 19: ...ead stopper Straighten infant s feet 3 Use left hand to secure infant s feet while using right hand to push foot stopper until it touches soles of infant s feet Press HOLD key to lock height measureme...

Страница 20: ...20 VIII Troubleshooting If head stopper comes loose please place it back in position...

Страница 21: ...ch Graduation 1 mm 1 16 in Accuracy 10 mm Dimensions Overall 870 W x 290 D mm Device Weight 0 6 kg Power Source AAA Batteries 2 Operation Temperature Humidity 5 35 Standard Accessories User manual x1...

Страница 22: ...________________________________________________ _______________________________________________________ _______________________________________________________ _______________________________________...

Страница 23: ...________________________________________________ _______________________________________________________ _______________________________________________________ _______________________________________...

Страница 24: ...ollowing directives 93 42 EEC as amended by 2007 47 EC Medical Device Directive Classification Class I with measuring function RoHS Directive 2011 65 EU and Delegated Directive EU 2015 863 Authorized...

Отзывы: