CGOLDENWALL NDJ-5S Скачать руководство пользователя страница 5

2

Installation

1

Take out base, lifting column, handle from the packing box. Screw

the column into the base ( rack side towards the user). Fix handle

onto the lifting block.

2

Adjust lifting button for tightness. Put T-shaped block on the rear

of the instrument into handle and get it fixed, making the instrument

balanced.

3

Adjust 3 screws on the base to keep the level bubble at the center

of the black circle.

4

Take off the protection cap at the bottom of the instrument.

5

Plug in.

5. Operating Procedure

1

Prepare liquid sample to be tested, pour it into beaker or flat

container with diameter no less than 60mm. Control the temperature

of the liquid sample correctly. (Variations of temperature will affect

viscosity directly.)

2

Screw anti-clockwise protection bracket (T-shape) into the bottom

end of the instrument

3

Choose appropriate rotor and screw it anti-clockwise into the

connector. Attention: slightly lift direction connector up when rotor is

installed to avoid damages to the pivot.

4

Turn lifting button to slowly immerse rotor into the liquid until the

marker on the rotor (groove or marking line) is aligned with the liquid

level.

5

Adjust the balance of the instrument again.

6

Should keep testing temperature stable to ensure accurate read out.

Otherwise readout may fluctuate.

6.

Panel Operation

Turn on power switch on the back of the instrument and the

instrument is on standby. The following information will be shown on

the panel.

S1

V6

T.....

%...

CP

000000

If Rotor 1 is used at 6RPM/min, can directly press the start

button on the panel to initiate testing. When display values become

stable, read the CP value. (In order to make readout stable, if viscosity

is relatively low, turn rotor about 3-5 rotations; if viscosity is

relatively high, turn rotor about 1-2 rotations.)

-

3

-

If what the user chooses is not the default value S1 and V6, press

rotor select and speed select button to enter selecting status, then press

confirm button to confirm. When set-up is done, press start button to

initiate testing.

Rotor S0~S4 and V0.1/V0.3/V0.6/V1.5/V3/V6/V12/V30/V60

will scroll displaying repeatedly.

Explanations:

S: stands for Rotor number. S1 refers to Rotor 1. (default value)

V:stands for rotation speed. V6 refers to 6RPM/min. (default

value)

T:.....

: Temperature

%........: test value’s percentage rate of the full value of the

measuring range.

Cp: mPa

·

S

Viscosity value

For example: If the viscosity of the liquid to be tested is estimated

3000mPa

·

S, it is recommended to choose the

combination S2

V6 or S3

V0.3

For the correlations between the combination of rotor and rotation

speed and the corresponding viscosity range, please refer to the below

table.

NDJ-1S,NDJ-5S measuring range table

1#    

2#    

3#    

4# 

1000   

5000   

20000   

100000

500   

2500   

10000   

50000

200   

1000   

4000   

20000

6

    

rotations

12

   

 rotations

30

   

 rotations

60

   

 rotations

100   

500   

2000   

10000

0#    

Rotating 

speed    

Rotor    

100   

50   
20 
10   

NDJ-8S,NDJ-9S measuring range table

1#    

2#    

3#    

4# 

1000   

5000   

20000   

100000

500   

2500   

10000   

50000

200   

1000   

4000   

20000

100   

500   

2000   

10000

0#    

100   

50   
20 
10   

6

     

 rotations

12

    

 rotations

30

    

 rotations

60

    

 rotations

0.3 

     

rotations

0.6 

    

rotations

1.5 

    

rotations

    

rotations

0.1 

     

rotations

200   

400   

1000   

2000   

6000   

2000  

10000  

40000  

200000

4000  

20000  

80000  

400000

10000  

50000  

200000  

1000000

20000  

100000  

400000  

2000000

60000  

300000  

1200000  

6000000

Rotating 

speed    

Rotor    

-

4

-

Отзывы: