Operatio
n’s manual
Rheometer MDR-01
14
A
B
C
Figure 13
shows the main display of the Configuration Torque
- Select A area to set Torque Standard unit for Calibrate
(dNm, lbin, and
kg-f).
- Select B area to set STD Torque value according to Correction
Torque Standard value for Calibrate.
- Select C area to set Warm up time (Torque Standard warming).
To enter or modify the values to any functions, proceed according to the
following steps;
1. Entering the parameters, and how to correct it
Select the required parameter >> enter the value >> Select “Enter” to
confirm.
2.
Using the “Clear” button
to clear the parameters
Select the required parameter >> Select “Clear” (the value will be reset
to
0)
.
3.
Using the “Back” button
To edit a numeric value, such as
100.0
to
125.0
, select the required
parameter field >> select “Back” to the required location >> enter the
value >> select “Enter” to confirm
.