Ceyear 4957D Скачать руководство пользователя страница 27

4957D/E/F Microwave Analyzer Quick Start Guide 

23 

press [OK] to complete the input. 

  Min 

Press  [Amplitude]  →  [Min  Scale],  enter  the  min  value  with  the  numeric  keys,  and 
press [OK] to complete the input. 

2)  Auto Scale 

With auto scale function of this mode, the display range of the measured power can be 
automatically adjusted in the following steps: 

Press [Amplitude] → [Auto Scale]. 

7. Setting the relative measurement function 

The relative measurement function reflects the changes of the power that has been set 
as  the  reference  signal.  The  readings of  the  power  value  changes are  expressed in dB 
and % respectively. When the relative measurement switch is turned on, the analyzer will 
measure and store the current power level, while the power measurement will display a 
power level relative to the saved value. Follow these steps: 

Press [Amplitude] → [Relative Off / On], and switch to [Relative Off / On]. 

When the relative measurement function is turned on, the power readings will be shown 
as 0 dB and 100%. For example, to measure a signal with a power of 5 dBm, when the 
relative measurement function is turned on and the signal power is reduced to 2 dBm, the 
power readings at this time are -3 dB and 50%. 

8. Setting the offset function 

When  the  power  of  the  measured  DUT  is  greater  than  the  maximum  power  that  the 
instrument  can  measure,  the  measured  power  can  be  reduced  within  the  normal 
measurement  range  by  connecting  an  attenuator.  The  power  offset  function  allows  for 
setting offset values to offset attenuation values or cable losses for added attenuators or 
connecting cables. It also allows for setting power offsets for gains of added amplifiers. 
Positive values can compensate losses and negative values can compensate gains. For 
example,  if  a  10  dB  attenuator  is  added  to  protect  the  probe,  to  avoid  excessive  input 
power, the offset value is set to 10 dB. At this time, the power reading of the analyzer is 
the actual reading. If an amplifier with a gain of 10dB is added to the measuring end, the 
offset  value  is  set  to  -10  dB,  and  then  the  power  reading  of  the  analyzer  is  the  actual 
reading. The function settings are as follows: 

Press [Amplitude] → [Offset Off / On] to enable the power offset function. 

9. Setting averaging 

The averaging function is generally used to measure low-power signals or signals close 
to  noise  power.  It  can  smooth  traces  to  reduce  the  influence  of  random  noise  on  the 
measurement  and  improve  measurement  accuracy,  but  at  the  same  time  it  will  reduce 
measuring speed. The times of averaging determine the times for reading averaging. The 
more  averaging  times,  the  more  the  noise  will  be  reduced.  The  times  of  averaging  are 
displayed  at  the  upper  left  corner  of  the  measurement  data  display  zone.  The  times  of 
averaging  can  be  set  between  1  ~  999.  Setting  the  averaging  times  will  not  affect  the 
calibration accuracy. The averaging function is set as follows: 

Press [BW] → [Ava Off / On] to turn on the averaging function, with the times set to 16 by 
default. You can also enter the desired averaging times with the numeric keys, and press 
[OK] to complete. 

3.7 Testing Skills 

The test 

steps of several test formats have been described in detail above, and now we’ll 

introduce some common skills during the measurement below. 

Содержание 4957D

Страница 1: ...4957D E F Microwave Analyzer Quick Start Guide China Electronics Technology Instruments Co Ltd...

Страница 2: ......

Страница 3: ...duct can meet the indicator requirement in the manual upon delivery The calibration and measurement are completed by measuring bodies with national qualification with relevant data to be provided for...

Страница 4: ...A Mode 9 3 2 DTF Measurement in the CAT Mode 15 3 3 Signal Measurement in the SA Mode 16 3 4 Measuring Signal Power in the Power Monitor PM Mode 18 3 5 Measuring the Electrical Length of Cables or Oth...

Страница 5: ...the following chapters and sections Get Prepared This chapter introduces the preparation before operation startup panel introduction and battery replacement of 4957D E F After reading this chapter you...

Страница 6: ...the instrument not mentioned in the manual Otherwise such consequences as reduced electromagnetic shielding property and internal component damage can occur affecting product reliability In this case...

Страница 7: ...e The basic parameters of the enclosed battery are as follows Nominal voltage 10 8 V Nominal capacity 7 800 mAh Working time About 2 5 h Charging time About 6 h Rechargeable batteries should not be ex...

Страница 8: ...s on the bottom of the touch screen are represented by XXX with XXX being the name of the key The menu keys on the right are represented by XXX with XXX being the menu name 2 3 Front Panel This sectio...

Страница 9: ...ch can be selected as CAT Network Analyzer Spectrum Analyzer Power Monitor USB Power Meter and Voltage Voltmeter etc Cal It is used to select the type of device under calibration piece perform calibra...

Страница 10: ...installed power connected b Battery installed and full power connected Flashing yellow a Battery installed and not full power connected Power on Normal on green a Battery not installed power connected...

Страница 11: ...arked as 10 MHz In Out with 50 impedance BNC female adapter 4957D E F microwave analyzers can receive externally input 10 MHz sine signals as phase locked reference signals and the input signals shoul...

Страница 12: ...ally switched from the speaker to the earphone 2 5 4 Instrument Symbols The instrument symbol yellow label shown in the figure indicates that the max power input from the test spectrum input port and...

Страница 13: ...g soft key to select the correct calibration kit model 5 Press OK return to the menu bar of Cal select the frequency response error correction function press M Cal Freq Resp OPEN S11 and the system wi...

Страница 14: ...pt Connect Thru cable between P1 and P2 then press corresp menu button to start Then connect N type cable between ports 1 and 2 All adapters or cables used in device measurement should be included i e...

Страница 15: ...location where the DUT is to be connected 5 After the display trace is stable press THRU and the analyzer will display Cal std measuring during the measurement of the standard piece and after completi...

Страница 16: ...etc 3 Open the calibration menu and press Cal 4 Press Cal Cal Kit to see if the calibration kit is consistent with the kit specified with the current soft key If not press the corresponding soft key t...

Страница 17: ...the breaker and connect the load to the test port 10 After the display trace is stable to measure the standard piece press LOAD the analyzer will display calibration standard measurement during the me...

Страница 18: ...e and connect the short circuit device to the test port 1 After the display trace is stable to measure the standard piece press P1 SHORT the analyzer will display calibration standard measurement duri...

Страница 19: ...layed in the information zone Connect the DUT for measure Figure 3 5 Standard piece connection of full 2 ports error correction In actual calibration there is no sequence for the open circuit device s...

Страница 20: ...w to display the submenu of Window and then select different window functions 5 Press Back Unit m ft to set the required distance unit system 6 Press Cable Specification and Load Cable to select the m...

Страница 21: ...ency value 4 Setting sweep span Press Freq Span Note whether the span data is displayed in the information input zone to determine the currently activated parameters Reduce the span to 20MHz for insta...

Страница 22: ...eed of connection to an external power probe can conveniently test the power of continuous wave signals with a certain bandwidth and the channel bandwidth can be set to a larger value so as to simulat...

Страница 23: ...onvenient operation interface as the vector voltmeter The results of single port electrical length measurement are displayed in amplitude and phase formats while the results of all port are only displ...

Страница 24: ...port error correction in the VVM mode 5 Selecting measurement mode The VVM mode of the analyzer allows for measurement of parameters of the DUT like amplitude phase VSWR and impedance etc 1 Press Mea...

Страница 25: ...USB Power Meter mode and press OK to complete Figure 3 11 Test interface of USB Power Meter mode 2 Connecting power probes 1 As shown in the figure connect the small end of the USB cable to 8723X ser...

Страница 26: ...required to set the frequency value of the measured CW Freq signal The USB power probe will store a set of correction factors for each frequency The frequency corresponding to the measured power signa...

Страница 27: ...e losses for added attenuators or connecting cables It also allows for setting power offsets for gains of added amplifiers Positive values can compensate losses and negative values can compensate gain...

Страница 28: ...F microwave analyzer provides two limit lines upper and lower When the amplitude of a signal in a certain frequency band exceeds the set upper limit or is less than the set lower limit the analyzer w...

Страница 29: ...rmat or csv data text format The Print Screen function allows for saving the information displayed on the current screen into the memory of the instrument in an image format The storage medium for sta...

Страница 30: ...environment for compliance with the requirements the test port connectors for any damages and the calibration piece for normal performance indicators If everything above is normal the instrument may b...

Страница 31: ...g to the following steps 1 Write a paper document describing the instrument failure and put it in the package together with the analyzer 2 Pack it with the original packing materials so as to minimize...

Страница 32: ...y 32 dB 40 dB 32 dB 40 dB 28 dB 35 dB Effective source matching 25 dB 37 dB 25 dB 37 dB 18 dB 30 dB Reflected tracking 0 10dB 0 14dB 0 10dB 0 25dB 0 10dB 0 25dB Transmission tracking 0 25dB 0 33dB 0 2...

Страница 33: ...nput level 27 dBm Test port type N type female 3 5 mm male 2 4 mm male Power supply Rechargeable li ion battery or power adapter Power consumption 45 W excluding battery charging Working temperature r...

Страница 34: ...a test software 4957D S03 Vector voltmeter software 4957D S04 USB power measurement software USB power probe to be equipped separately 4957D S05 Power meter software 4957D S06 Field strength measureme...

Страница 35: ...female male calibration cable GORE 0K0CJ0CK024 0 4957D H14 87230 USB CW power probe 4957D H15 87231 USB CW power probe 4957D H16 87232 USB CW power probe 4957D H17 87233 USB CW power probe 4957D H18...

Страница 36: ...4957D E F Microwave Analyzer Quick Start Guide 32 4957D H32 Soft backpack 4957D H33 Aluminum alloy carrying case 4957D H34 Waterproof case...

Отзывы: