3 SCPI
3.3 Instrument-specific Commands
65
3.3.7.3 Corrected Measurement Results in Fixed Frequency Mode
[ Noise Figure Measurement - Corrected ]
:FETCH:SCALar:CORRected:NFIGure?
Function description:
This command is used to query the calibrated noise figure value at a fixed
frequency point.
Query format:
:FETCH:SCALar:CORRected:NFIGure? DB|LINear
Parameter description:
The unit is DB or LINear.
If no unit is specified, the system default unit dB is used.
Return value:
Noise figure value
Example:
:FETCH:SCALar:CORRected:NFIG? DB
[ Gain Measurement - Corrected ]
:FETCH:SCALar:CORRected:GAIN?
Function description:
This command is used to query the calibrated gain value at a fixed frequency point.
Query format:
:FETCH:SCALar:CORRected:GAIN? DB|LINear
Parameter description:
The unit is DB or LINear.
If no unit is specified, the system default unit dB is used.
Return value:
Gain
Example:
:FETCH:SCALar:CORRected:GAIN? DB
[ Y Factor Measurement - Corrected ]
:FETCH:SCALar:CORRected:YFACtor?
Function description:
This command is used to query the calibrated Y factor at a fixed frequency point.
Query format:
:FETCH:SCALar:CORRected:YFACtor? DB|LINear
Parameter description:
The unit is DB or LINear.
If no unit is specified, the system default unit dB is used.
Return value:
Y factor
Example:
:FETCH:SCALar:CORRected:YFAC? DB
[ Cold Power Measurement - Corrected ]
:FETCH:SCALar:CORRected:PCOLd?
Function description:
This command is used to query the calibrated cold power value at a fixed frequency point.
Query format:
:FETCH:SCALar:CORRected:PCOLd? DB|LINear
Parameter description:
The unit is DB or LINear.
If no unit is specified, the system default unit dB is used.
Return value:
Cold power value