3 SCPI
3.3 Instrument-specific Commands
35
Reset status:
4Mhz
Key path:
Front panel
【
BW
】
—>[Man]
[ Bandwidth Mode ]
:SENSe:NFIGure:BANDwidth:AUTO
Function description:
This command is used to set/query manual or auto mode of bandwidth.
Setting format:
:SENSe:NFIGure:BANDwidth:AUTO ON|OFF|0|1
Query format:
:SENSe:NFIGure:BANDwidth:AUTO?
Parameter description:
ON|1: auto mode
OFF|0: manual mode
Example:
:SENSe:NFIGure:BANDwidth:AUTO ON
:SENSe:NFIGure:BANDwidth:AUTO?
Reset status:
1
Key path:
Front panel
【
BW
】
3.3.3 ENR
The excess noise ratio corresponds to the
【
ENR
】
key on the front panel and is used to set the excess
noise ratio and edit the excess noise ratio data. These include setting ENR mode, editing excess noise
ratio table, specifying a cold temperature, specifying a fixed cold temperature, selecting a fixed
frequency excess noise ratio, and setting SNS.
[ ENR Mode ]
:SENSe:CORRection:ENR:MODE
Function description:
This command is used to set/query the ENR mode.
Setting format:
:SENSe:CORRection:ENR:MODE TABLe|SPOT
Query format:
:SENSe:CORRection:ENR:MODE?
Parameter description:
TABLe Table
SPOT Fixed
Example:
:SENS:CORR:ENR:MODE TABL
:SENS:CORR:ENR:MODE?
Reset status:
Table
Key path:
Front panel
【
ENR
】
—>[ENR Mode]
[ Common Table ]
:SENSe:CORRection:ENR:COMMon:STATe
Function description:
This command is used to set/query the common table status.
Setting format:
:SENSe:CORRection:ENR:COMMon:STATe OFF|ON|0|1
Query format:
:SENSe:CORRection:ENR:COMMon:STATe?
Parameter description:
ON|1: The measurement ENR table is used for measurement and calibration
when the status is enabled.