![Ceyear 1465 Series Скачать руководство пользователя страница 128](http://html1.mh-extra.com/html/ceyear/1465-series/1465-series_programming-manual_2584913128.webp)
5 Error Description
5.1 Error information
121
5 Error Description
This chapter aims to help you to identify a problem and receive after-sales service. Error information of
the signal generator is also introduced.
Error information ................................................................................................ 121
How to fix errors ................................................................................................ 123
5.1 Error information
The signal generator uses two methods to record the errors during measurement: the front panel
operation interface displays the error message queue and the SCPI (remote control mode) error
message queue, which are separately stored and managed.
Description of error messages .......................................................................... 121
5.1.1 Description of error information
Local error information ...................................................................................... 121
Remote error information .................................................................................. 122
5.1.1.1 Local error information
Error information view ....................................................................................... 121
Description of error messages .......................................................................... 121
1) Error information view
Operation method through interface:
If any error information is displayed in the lower right corner of the signal generator during use, it
indicates that the operation of the software or hardware of the instrument is defective. You can
determine the type of the fault according to the error code and then take appropriate measures for
troubleshooting.
The error display area of the signal generator can only display one piece of error information at one time.
As the instrument may have a number of problems at the same time, all error information can be seen by
executing the following operations:
Step 1.
Press
【
System
】
, then press [Error List]. A window appears with a list of errors.
Step 2.
The prompt information is displayed in the window.
Step 3.
Use the mouse to browse the error information and close the dialog window.
Step 4.
Select the Clear List button to clear the history error information.
2) Description of error messages
If errors are detected during measurement by the signal generator, the warning or error information
(error abbreviation + detailed description of error) will be displayed on the right side of the status
indication area.
Table 5.2 Local Error Information List
Error Key Field
Description
Data error
Only for digital-based band interface: Digital I/Q input data error.