![Ceyear 1433D Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/ceyear/1433d/1433d_quick-start-manual_2584928005.webp)
4
Technical Specification
1.
Main Functions
CW signal output function
FM/AM/Pulse modulation function
Step and List sweep function
Unstable amplitude and Unlock alarm function
Support LAN and USB interface program control function
2.
Specifications
Frequency Characteristics
Frequency Range
1433D:1MHz
~
20GHz
1433E:1MHz
~
26.5GHz
1433F:1MHz
~
40GHz
1433H:1MHz
~
50GHz
Frequency
N
(
internal YO
harmonic number
)
1MHz
≤f
<
2.35GHz
1/2
2.35GHz≤f
<
2.5GHz
1/8
2.5GHz
≤f
<
5GHz
1/4
5GHz≤f≤10GHz
1/2
10GHz
<
f≤20GHz
1
20GHz
<
f≤40GHz
2
40GHz
<
f≤50GHz
4
Frequency Resolution
0.1Hz
Internal timebase
Aging rate
± 0.5× 10
-6
/year
Temperature effects
± 0.3× 10
-6
(
-10
℃~
50
℃,
versus 25
℃
± 5
℃)
Initial calibration accuracy
± 0.5× 10
-6
Reference input
frequency
1MHz
~
10MHz
,
step 1MHz
Power
-5dBm
~
+10dBm
,
impedance: 50
Ω
Reference output
Frequency
10MHz
Power
>
0dBm
,
50
Ω
impedance
Sweep characteristics
Sweep features
Sweep mode
Step/List
Sweep points
2 to 1601
Dwell time
10ms to 100s
Trigger mode
Auto/manual
Level characteristics
Stable output power
range(25
℃
± 10
℃,
CW
mode)
1MHz≤f
<
2.5GHz
-120dBm
~
+5dBm
2.5GHz≤f≤10GHz
-120dBm
~
+10dBm
10GHz
<
f≤20GHz
-120dBm
~
+5dBm
20GHz
<
f≤40GHz
-120dBm
~
+5dBm