4.

 

When in position, exert additional forward force in line with the shaft prior to and during needle 

deployment. Slowly deploy the needle by simultaneously squeezing Levers A and B (see Figure 1 

above) until Lever B is fully advanced.  Forward pressure on the device prior to and during 

deployment will prevent the tissue from migrating distally. Distal tissue migration during 

deployment may lead to suture misfire.  

 

5.

 

During deployment, the suture outside of the access port should be observed advancing in tandem 

with the needle-trigger advancement. If the suture does not advance, or if the suture breaks, the 

suture may be pinched between the device and cannula seal. If this occurs, after reloading the device 

with suture consider advancing the device beyond the tear in order to introduce slack in the suture 

prior to positioning within the tear.  

 

6.

 

Release both Levers A and B and then retract the lower jaw by pushing down Lever C. Engage Lever 

A and remove the device. Keep Lever A engaged when withdrawing the device from the joint in 

order to keep the upper jaw parallel with the shaft.   Remove the suture end from the upper jaw by 

pulling in the proximal direction, parallel to the upper jaw. 

 

7.

 

If the 2-0 or 0 suture is being passed to subsequently shuttle #2 suture, create a loop in the suture 

limb that exits on the glenoid side of the labral tissue and use that loop to shuttle the desired suture 

through the labral tissue.  If a loop is created prior to loading the suture, the free (non-looped) end 

should be loaded into the device.  This will ensure that the loop is positioned on the glenoid side of 

the tear.   

 

8.

 

Complete the procedure as usual. 

 

GENERAL USE 

 

Bench deployment testing has demonstrated that the device can be reloaded with suture and fully 

actuated 14 times. Cycling the device in excess of 14 deployments may impact needle integrity and 

compromise performance. 

 

Immediately discard the Ceterix NovoStitch Disposable Suture Passer after completion of the surgery.  

The device is 

SINGLE USE ONLY

 

 

STORAGE 

Store in a cool, dry area. 

 

P/N:  LBL-0012     Rev: E 

 

 

Page  6 

 