QUICK GUIDE CB011/CB012
INCLUDES
G_CB011-CB012_v0003_20211013_EN
Manual
(22 pages)
available at:
https://www.cesva.com/en/support/product/
CB011
CB012
2 AA Alkaline
batteries
Calibrator
CB011 o CB012
POWER AND TURN ON/OFF THE CALIBRATOR
Open the battery compartment
with a screwdriver and insert the
3 batteries respecting the polarity
To turn the calibrator ON /
OFF press
To keep the calibrator on perma-
nently, press
for few seconds
CHECKING THE DEVICE
1º Insert the device into the calibrator (see figure)
2º Turn on the device*
3º Turn on the calibrator
4º Start measuring with the device*
5º Check that the function value on the screen coincides with the value
If you work with microphones the value corrected is 93,9 dB
6º If the difference between the values is within the recommended ranges*, the
verification is favorable. Otherwise, contact the nearest
official
service.
7º Turn off the calibrator and the device
* Check the device
manual.