Cerlic ITX 20 Скачать руководство пользователя страница 9

                                                                                                        

 

 

- 9 - 

9.

 

Menu for ITX sensor 

Use 

 or 

 to select the sensor in the main display. The menu for the selected sensor 

is accessed by pressing ENTER for five seconds. If the selected sensor is not active (the 
text 

No transmitter

 is shown) a warning is displayed that asks you to make another 

choice in order to show the sensor menu. 

 

Settings 

I-Time

 

Integration time or dampening - can be set up to 999 seconds 

Unit 

”%”, ”ppm”, ”g/l”, or ”mg/l”

 

Alarm Relay  

”-” ”#1”, ”#2”, or ”#1 and #2”. Check that the relay is not being 
used for cleaning 

Analog 

 

”None”, ”Channel 1”, ”Channel 2”, or ”Both”. Pick which analog 
output contacts are to be used with sensor 

Cleaning 

Press ENTER to go to Cleaning program 

Cleaner  

”None”, ”Flush”, or ”Brush”. Do not select “Brush” since this 
does not exist for ITX sensor

 

Interval min 

0-999 minutes, time between cleaning cycles

 

Length sec 

0-999 seconds, duration of flushing cycle

 

Freeze sec 

0-999 seconds, extra freeze time of output signal after a 
flushing cycle

 

 

Relay 

”-”, ”#1”, or ”#2”. Select relay to operate solenoid for flush cycle. 
These are the same relays used for “Alarm relay” above

 

 

 

Next time  

The next scheduled cleaning time. Pushing “Enter” on this line 
will set the time to current time and start a cleaning cycle. This 
could be used to test the “Flush” cycle

Calibrate 

Take sample 

No, Zero, #1, n#2, #3, #4, #5. Sensor stores current MS (light) 
value in memory and you enter a lab solids value below to 
complete calibration. 

Con 

Current concentration, (same as shown in the main menu)  

Sample #1 

Lab test – consistency/suspended solids value for Sample 1 

Sample #2 

Lab test – consistency/suspended solids value for Sample 2 

Sample #3 

Lab test – consistency/suspended solids value for Sample 3 

Sample #4 

Lab test – consistency/suspended solids value for Sample 4 

Sample #5 

Lab test – consistency/suspended solids value for Sample 5 

Содержание ITX 20

Страница 1: ...2002 11 19 MANE11 02 Ver 2 00 ITX 20 Suspended Solids Sensor...

Страница 2: ...g plate for cleaning solenoid valves 8 9 Menu for ITX sensor 9 Settings 9 Calibrate 9 System 10 10 Calibration 11 Getting good measurement 11 Calibration screen 11 Zero Calibration 12 Entering sample...

Страница 3: ...ntenance under critical applications The measuring lenses within the housing are made of glass The electronics and optics are protected in the rugged casing ensuring its reliability in very demanding...

Страница 4: ...check that the content corresponds to your order and packing list Every shipment should include ITX Sensor w 10 m 33ft cable and flushing hose P N 11305455 Optional parts for ITX Suspended Solids Sens...

Страница 5: ...lation Tips Adjust the rod so that the sensor is at least 12 below the liquid surface or the lowest water level in decant applications to prevent the sensor from coming out of the liquid In an aeratio...

Страница 6: ...re all water inside the rod is drained Clean the sensor and rod with a brush or clean cloth Do not use a wire brush Open the black sensor adapter Flush the inside of the rod with clean water Mount the...

Страница 7: ...e the control box For example if the solenoid is wired to relay 1 select 1 for flushing Next time displays the next time flush will be activated pushing ENTER will set it to current time and those sta...

Страница 8: ...in flushing nozzles Cerlic offers a predrilled Aluminum handrail mounting plate The Mounting plate is fixed to the rail using two u bolts Please see technical specifications for details and ordering i...

Страница 9: ...ing cycles Length sec 0 999 seconds duration of flushing cycle Freeze sec 0 999 seconds extra freeze time of output signal after a flushing cycle Relay 1 or 2 Select relay to operate solenoid for flus...

Страница 10: ...e SA 1 SA value for sample 1 Cons 1 Lab test consistency or suspended solids value for Sample 1 SA and Cons repeated for samples 2 to 5 Info Press ENTER to go to info read only menu This menu is for C...

Страница 11: ...es while new calibrations will start from scratch Calibration screen The ITX sensor information menu is the calibration curve screen To change between the main menu and the calibration screen press an...

Страница 12: ...ect unit consistency suspended solids is selected for the application Select unit in the sensor menu SETTINGS UNIT Remove the sensor from the process clean the sensor head and dip it in a bucket of cl...

Страница 13: ...et so the sensor is at least 12 36 cm below the lowest liquid level in the basin Take the bucket to the lab for analysis Note the concentration of the sample determined at the lab Enter sample 1 conce...

Страница 14: ...to set high and low alarms values that can be used to notify when solids have reached critical points MAX sets the 20 mA point output MIN sets the 4 mA point output may be a negative value for specia...

Страница 15: ...loped and manufactured in accordance with the following standards SS EN50 082 2 immunity SS EN50 081 2 emission SS EN61 010 1 safety 89 336 EEC 92 31 EEC 93 36 EEC 73 23 EEC Optional parts for ITX Sus...

Страница 16: ...16 13 Dimensions 2 60 8 94 10 04 2 60 66...

Страница 17: ...cket Make sure that the bracket guide tracks are properly seated in the bracket Fasten the safety locking clamp Check that the bracket is safely fixed to the rail for the spring to work the way it is...

Страница 18: ...18 Appendix 2 Assembly of adjustable slide rail...

Страница 19: ...me Use or to select the sensor in the main display Go to the sensor menu it is accessed by pressing ENTER for five seconds Select System and press ENTER Type __________________________________________...

Страница 20: ...___________________________________________ Cons 1 _____________________________________________________ Ch1a _____________________________________________________ Ch1 ________________________________...

Страница 21: ...________________________ Alarm Relay _____________________________________________________ Analog _____________________________________________________ Cleaner ________________________________________...

Отзывы: