
- 9 -
9.
Menu for ITX sensor
Use
or
to select the sensor in the main display. The menu for the selected sensor
is accessed by pressing ENTER for five seconds. If the selected sensor is not active (the
text
No transmitter
is shown) a warning is displayed that asks you to make another
choice in order to show the sensor menu.
Settings
I-Time
Integration time or dampening - can be set up to 999 seconds
Unit
”%”, ”ppm”, ”g/l”, or ”mg/l”
Alarm Relay
”-” ”#1”, ”#2”, or ”#1 and #2”. Check that the relay is not being
used for cleaning
Analog
”None”, ”Channel 1”, ”Channel 2”, or ”Both”. Pick which analog
output contacts are to be used with sensor
Cleaning
Press ENTER to go to Cleaning program
Cleaner
”None”, ”Flush”, or ”Brush”. Do not select “Brush” since this
does not exist for ITX sensor
Interval min
0-999 minutes, time between cleaning cycles
Length sec
0-999 seconds, duration of flushing cycle
Freeze sec
0-999 seconds, extra freeze time of output signal after a
flushing cycle
Relay
”-”, ”#1”, or ”#2”. Select relay to operate solenoid for flush cycle.
These are the same relays used for “Alarm relay” above
Next time
The next scheduled cleaning time. Pushing “Enter” on this line
will set the time to current time and start a cleaning cycle. This
could be used to test the “Flush” cycle
.
Calibrate
Take sample
No, Zero, #1, n#2, #3, #4, #5. Sensor stores current MS (light)
value in memory and you enter a lab solids value below to
complete calibration.
Con
Current concentration, (same as shown in the main menu)
Sample #1
Lab test – consistency/suspended solids value for Sample 1
Sample #2
Lab test – consistency/suspended solids value for Sample 2
Sample #3
Lab test – consistency/suspended solids value for Sample 3
Sample #4
Lab test – consistency/suspended solids value for Sample 4
Sample #5
Lab test – consistency/suspended solids value for Sample 5