Cerlic 11305473 Скачать руководство пользователя страница 8

 

 

- 8 - 

 

9.

 

Cleaning unit 

The transmitter is equipped with built- in flushing nozzles.  The nozzles are used to 
direct the cleaning medium, compressed air or water, via a flushing hose that is 
connected to the top of the transmitter. The air or liquid is controlled with a solenoid 
valve that is wired into the BB2 control box. 

Flushing must be activated in the “Settings” menu in the BB2 control box: 

 

Select the O2X transmitter in the Main Menu by using the 

 or 

 buttons. 

 

Press ENTER for approximately five seconds to enter the Transmitter Menu. 

 

Use the arrows to select “Settings” and press ENTER. 

 

Use the arrows to select “Cleaning” and press ENTER. 

 

In the “Cleaning” submenu, adjust the “Cleaner” setting to “flush” and specify 
the cleaning interval in minutes and the flush time in seconds. 

 

Specify the relay to be used according to the wiring inside the control box.  For 
example, if the solenoid is wired to relay #1, select “#1” for flushing. 

 

“Next time” displays the next time flush will be activated, pushing ENTER will 
set it to current time and those start cleaning.. 

 

 

 

 

 

 

 

Cleaning the Flushing Nozzle  

If the flushing nozzle becomes plugged, removing the flushing head to backflush the 
nozzles with clean water can clean it.  Before removing the flushing head, close the 
valve for the flush water source. Place a ½ ” hose over the flush nozzle and carefully 
open the water valve.  The pressure should clear the line of solids.  If backflushing does 
not work initially, try cleaning the three flushing nozzles with a  needle.  Try 
backflushing the nozzles again as described above until clean water comes out. 

Flushing tube 

Flushing 

water 

Sensor cable 

Note! 

 

Pay attention to the requirements for protection against backflow, according to 
the EN 1717 standard for drinking water devices.  If possible, use plant reuse 
water or effluent water for cleaning. 

 

Never flush if the electrode is not installed. 

 

When the electrode is not installed, the yellow plastic cap should be kept on the 

Содержание 11305473

Страница 1: ...2002 11 18 MANE11 02 Ver 2 00 O2X Dissolved Oxygen Sensor...

Страница 2: ...Installing a new electrode 7 Mounting the protection plate 7 9 Cleaning unit 8 Cleaning the Flushing Nozzle 8 Mounting plate for solenoid valves 9 10 Menu for O2X Transmitter 10 11 Calibration 12 Zer...

Страница 3: ...t in flushing nozzles allow for the most accurate readings with little maintenance The electronics is protected in the rugged casing ensuring its reliability in very demanding environments The sensor...

Страница 4: ...hould include O2X sensor P N 11305473 including 33 ft 10 m cable and 33 ft 10m flushing hose You will also need an electrode below to get the sensor working Optional parts for O2X Dissolved Oxygen sen...

Страница 5: ...liquid surface or the lowest water level in decant applications SBRs to prevent the sensor from coming out of the liquid In an aeration tank ensure that the sensor is not directly above a diffuser he...

Страница 6: ...ing a new electrode The electrode can usually be both assembled and disassembled without tools Do not ever clamp onto the sensor housing or part of the cable when installing or removing the flushing h...

Страница 7: ...nting Tighten the flushing head and mount the protection plate if needed See below for instructions on mounting the protection plate A new air calibration must be performed whenever the electrode is c...

Страница 8: ...olenoid is wired to relay 1 select 1 for flushing Next time displays the next time flush will be activated pushing ENTER will set it to current time and those start cleaning Cleaning the Flushing Nozz...

Страница 9: ...lushing nozzles Cerlic offers a predrilled Aluminum handrail mounting plate The Mounting plate is fixed to the rail using two u bolts Please see technical specifications for details and ordering infor...

Страница 10: ...to be seen as additional information not a primary measurement Cleaning Press ENTER to go to Cleaning program Cleaner None Flush or Brush Do not select Brush since this does not exist for O2X sensor I...

Страница 11: ...ly SoftW Software version of the sensor read only Temp Sensor temperature read only Info Press ENTER to go to info read only menu This menu is for Cerlic internal use it may change without notice MS0...

Страница 12: ...brating a sensor Air calibrating a sensor The most important thing to keep in mind while performing an air calibration is that it is necessary to create a constant environment The electrode cartridge...

Страница 13: ...ar hPa 29 2 990 29 3 993 29 4 997 29 5 1000 29 6 1003 29 7 1007 29 8 1010 29 9 1014 30 0 1017 30 1 1020 30 2 1024 30 3 1027 30 4 1030 The following appear in the Calibration menu as read only Temp inf...

Страница 14: ...to set high and low alarms values that can be used to notify when the level has reached critical points MAX sets the 20 mA point output MIN sets the 4 mA point output may be a negative value for speci...

Страница 15: ...SS EN50 082 2 immunity SS EN50 081 2 emission SS EN61 010 1 safety 89 336 EEC 92 31 EEC 93 36 EEC 73 23 EEC Optional parts for O2X Dissolved Oxygen Sensor Electrode cartridge P N 11805160 Mounting br...

Страница 16: ...16 14 Dimensions...

Страница 17: ...cket Make sure that the bracket guide tracks are properly seated in the bracket Fasten the safety locking clamp Check that the bracket is safely fixed to the rail for the spring to work the way it is...

Страница 18: ...18 Appendix 2 Assembly of adjustable slide rail...

Страница 19: ...____________ Box heat _____________________________________________________ Leave the BB2 menu by pressing and ENTER at the same time Use or to select the sensor in the main display Go to the sensor m...

Страница 20: ...Relay _____________________________________________________ Analog _____________________________________________________ Cleaner _____________________________________________________ Cleaning interva...

Отзывы: