42

2. Attached documents

The CGM MCB-1201 is intended for use in the electromagnetic environment

specified below. The customer or the user of CGM MCB-1201 should assure

that it is used is such an environment.

Immunity test

Compliance level

IEC 60601

Test level

Electromagnetic

environment - guidance

Electrostatic

discharge

(ESD)

IEC 61000-4-2

6kV Contact

8kV Air

6kV Contact

8kV Air

Floors should be wood,

concrete or ceramic tile.

If floors are covered with

synthetic material, the

relative humidity should be

at least 30%. 

Electrical fast

Transient/burst

IEC 61000-4-4

2kV for power

supply lines 1kV

for input/output

lines

2kV for power

supply lines 1kV

for input/output

lines

The main power’s quality 

should be that of a typical

commercial or hospital

environment.

Surge

IEC 61000-4-5

1kV differential

mode

2kV common

mode

1kV differential

mode

2kV common

mode

The main power’s quality 

should be that of a typical

commercial or hospital

environment.

Power frequency

(50/60Hz)

Magnetic field

IEC 61000-4-8

3.0 A/m

3.0 A/m

Power frequency magnetic

fields should be at levels

characteristic of a typical

location in a typical

commercial or hospital

environment.

Note : 

U

т is the a.c. mains voltage prior to application of the test level.

<5% 

U

т 

(>95%

dip in 

U

т)

for 0.5 cycle

40% 

U

т (60% dip

in 

U

т) for 5 cycle

70% 

U

т (30% dip

in 

U

т)for 25 cycle

<5% 

U

т 

(<95%

dip in 

U

т)

for 5 s

<5% 

U

т 

(>95%

dip in 

U

т)

for 0.5 cycle

40% 

U

т (60% dip

in 

U

т) for 5 cycle

70% 

U

т (30% dip

in 

U

т)for 25 cycle

<5% 

U

т 

(<95%

dip in 

U

т)

for 5 s

The main power’s quality 

should be that of a typical

commercial or hospital

environment. If the user of

the CERAGEM-CH requires

continued operation during

power mains interruptions,

it is recommended that the

CGM MCB-1201 

be powered

from an uninterruptible

power supply or a battery.

Manufacturer’s declaration - electromagnetic immunity

43

2. Attached documents

The 

 

is intended for use in the electromagnetic environment specified 

below. 

The customer or the user of

  the 

  should assure that it is used in such an environment.

 

Immunity 

test 

IEC 60601 

Test level 

Compliance 

level 

Electromagnetic environment -

guidance 

Conducted 

RF 

IEC 61000-4-

6

 

3 Vrms 

150 kHz to 80 

MHz

 

3 Vrms 

150 kHz to 80 

MHz

 

Portable and mobile RF communications 

equipment should be used no closer to any part 
of the 

, including cables, than 

the recommended separation distance calculated 

from the equation applicable to the frequency of 

the transmitter. 

 

Recommended separation distance 

 

 

Radiated RF 

IEC 61000-4-

 

3 V/m 

80.0 MHz to 2.5 

GHz 

3 V/m 

80.0 MHz to 2.5 

GHz 

Recommended separation 

distance 

 

 

 

Where 

is the maximum output power rating of 

the transmitter in watts (W) according to the 

transmitter manufacturer and 

is the 

recommended separation distance in meters (m). 

Field strengths from fixed RF transmitters, as 

deter-mined by an electromagnetic site survey,   

(a) Should be less than the compliance level in 

each frequency range (b). 

 

Interference may occur in the vicinity of 

equipment marked with the following symbol:

 

 

Note 1)

 

U

т

 is the A.C. mains voltage prior to application of the test level. 

 

Note 2)

 At 80 MHz and 800 MHz, the higher frequency range applies. 

 

Note 3)

 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. 

To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be 

considered. If the measured field strength in the location in which the EUT is used exceeds the applicable RF 

compliance level above, the EUT should be observed to verifynormal operation. If abnormal performance is observed, 

additional measures may be necessary, such as re-orienting or relocating the EUT. 

 

b

 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V / m.

 

CGM MCB-1201

CGM MCB-1201

CGM MCB-1201

Содержание CGM MCB-1201

Страница 1: ......

Страница 2: ...is possibility of minor injury or property damage 1 While using this product do not use pins or other metals 2 Do not use the product when there is humidity or water 3 The external projector must be u...

Страница 3: ...rs 8 Do not use the product where there are foreign substances in the air or places that handle oxygen 9 Do not use or apply skin care cosmetic or medical products on your skin before and during the u...

Страница 4: ...sk of falling 9 Use according to the instruction manual and keep to the duration and methods instructed in the manual Use 1 2 times a day and do not use other products unless supplied by the manufactu...

Страница 5: ...conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation NOTE This equipment has be...

Страница 6: ...ion Installation and preparation 11 1 Remote control layout 2 How to use the remote control 3 How to use the product by mode 4 How to move the product Using the product 17 1 Self inspection 2 Manageme...

Страница 7: ...6 MEMO...

Страница 8: ...7 Introduction of the product 1 Product s characteristics 2 The name of each part...

Страница 9: ...ive medicine allow users to select programs that b needs 5 Slim Easy To Use Remote Control The slim and light remote control has function keys in the form of icons for more convenient product operatio...

Страница 10: ...ame Main Heating Element Remote control Internal Projector Sensor for the opening of auxiliary part detection Sensor for the opening of auxiliary part detection It detects that the auxiliary part is o...

Страница 11: ...10 MEMO...

Страница 12: ...11 Installation and preparation 1 Check product s composition 2 Check before product installation 3 Installation 4 Check after installation...

Страница 13: ...r Remote control Simple to operate and check the product s status Remote control back label Describes the remote control button layout and can be attached to the back of the remote control Power cord...

Страница 14: ...ll according to the designated order 5 Do not drop the product or apply any force to it This could cause malfunctioning 4 When installing check that the rated voltage is 100 240V To avoid circuit dama...

Страница 15: ...g the supporting frame Take care not to insert any part of your body into the product when opening the product There is risk of injury 3 2 Lay the product where the ground is flat and unlock the safet...

Страница 16: ...erly 4 Remote Control External Projector If the power cord is connected when the power switch is On it can cause mechanical failure Caution Connect the power cord onto the main body and plug into a so...

Страница 17: ...efore connecting the power cord connect the remote control first 2 Set the temperature appropriately according to user when using the product 3 After using the product turn off power on the remote con...

Страница 18: ...1 Remote control layout 2 How to use the remote control 3 How to use the product by mode 4 How to move the product Using the Product 17...

Страница 19: ...tion Power button Semi automatic mode symbol Manual mode symbol Internal projector position symbol Automatic mode button Mode A Internal projector s temperature symbol External projector s temperature...

Страница 20: ...3 seconds to turn it off Start the remote control after opening the auxiliary part 2 Check the display on the remote control External projectors appear only when they are connected The current tempera...

Страница 21: ...s completed 3 Check the changed temperature set point on the remote control display If you do not press the setting button after setting the temperature by pressing the internal external projectors te...

Страница 22: ...state after pressing the mode button 2 Select the mode by pressing the automatic mode button Mode A 3 The mode begins automatically and body sensing starts Lie in the correct posture When you move th...

Страница 23: ...Select the mode by pressing the automatic mode 1 button 3 The mode begins automatically and body sensing starts Lie in the correct posture When you move the internal projector the moving projector dis...

Страница 24: ...tor display appears Automatic Mode Mode 2 3 How to use the product by mode 23 1 Turn on the remote control power and check the display 3 The mode begins automatically and body sensing starts Lie in th...

Страница 25: ...the mode number 2 Select the mode by pressing the mode button Modes 3 9 3 The mode begins automatically and body sensing starts Lie in the correct posture When you move the internal projector the mov...

Страница 26: ...more in 5 second standby state after pressing the mode button 2 Press the semi automatic button and move the internal projectors to the desired position 3 The mode will automatically begin centered on...

Страница 27: ...the mode will then be changed immediately 2 Press the manual mode button and move the internal projectors to the desired position 3 Thermal stimulation begins from the designated position Assume a pro...

Страница 28: ...3 Scanning Thermal acupressure 88 seconds 2 minutes Standby Abdomen switch 3 5 7 4 12 10 8 6 4 Cervical spine 1 Sacral 3 Sacral 1 Lumbar 2 Thoracic vertebrae 2 10 seconds Standby 3 5 7 4 12 10 8 6 4...

Страница 29: ...e 5 Mode 6 Mode 7 3 5 7 4 12 10 8 6 4 10 seconds Standby 3 5 7 4 12 10 8 6 4 Cervical spine 1 Sacral 3 Sacral 1 Lumbar 2 Thoracic vertebrae 2 10 seconds Standby Cervical spine 1 Sacral 3 Sacral 1 Lumb...

Страница 30: ...1 Lumbar 2 Thoracic vertebrae 2 10 seconds Standby Cervical spine 1 Sacral 3 Sacral 1 Lumbar 2 Thoracic vertebrae 2 Mode Pattern Scanning Thermal acupressure 88 seconds Mode Duration Approx 18 minute...

Страница 31: ...wn There is risk of injury and equipment failure Caution Take care not to insert any part of your body into the product when closing the product There is risk of injury Hold the handle and move the pr...

Страница 32: ...Self inspection 2 Managementtechniquesandstoragemethods 3 Check before A S request 4 Remote control error messages 5 Product specifications 6 Product warranty Management and matters that require atten...

Страница 33: ...ector s temperature has been set to the appropriate level 5 Is there anything strange on the exterior 6 Did you check whether there is no damage to the external projector cover 1 Did you check whether...

Страница 34: ...n wipe with a dry cloth 3 If there is contact with cosmetic products or volatile emulsions and is left for a long time the surface may peel off so wipe with a dry cloth as soon as possible 4 When wash...

Страница 35: ...ch at the bottom of the product turned on A Turn on the power switch Q Is the power cord properly connected A Reconnect the power cord Q Is the remote control properly connected A Reconnect the remote...

Страница 36: ...n the external projector s temperature is over 70 the E2 error appears on the temperature display of the external projector and all operations of the product stop If the motor is overloaded and certai...

Страница 37: ...olded 1170mm x 600mm x 175mm 5mm To improve quality design and product specifications may be changed without notice 26kg 2 Environment temperature 10 40 Purpose of use Maximum Permissible Weight 135kg...

Страница 38: ...licable at the time of purchase or contract This letter of warranty is valid only to the person who purchased the product and therefore those who acquired this product by transfer from the initial pur...

Страница 39: ...38 MEMO...

Страница 40: ...1 Explanation of mark delivery and storage 2 Attached documents Service manual 39...

Страница 41: ...he marked area to exchange the fuse related to service engineer Capacity of a fuse AC 250V T2AL Time Lag 2A Glass Body Fuse TYPE BF APPLIED PART V AC alternating current voltage source On power connec...

Страница 42: ...netic environment guidance RF emissions CISPR 11 Group 1 The Model CGM MCB 1201 uses RF energy only for its internal function Therefore Its RF emissions are very low and are not likely to cause any in...

Страница 43: ...tial mode 2kV common mode 1kV differential mode 2kV common mode The main power s quality should be that of a typical commercial or hospital environment Power frequency 50 60Hz Magnetic field IEC 61000...

Страница 44: ...n the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note 1 U is the A C mains voltage prior to application of the test...

Страница 45: ...below according to the maximum output power of the communications equipment Rated maximum output power W of transmitter Separation distance m according to frequency of transmitter 150 kHz to 80 MHz 8...

Страница 46: ...t marked with the following symbol Note 1 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Note...

Страница 47: ...46 MEMO...

Страница 48: ...47 MEMO...

Страница 49: ...48 MEMO...

Страница 50: ......

Отзывы: