Figure 2:

Typical

Gain Performance

Figure 3:

Typical

Reflection

Performance

Figure 4:

Typical

Group Delay

Performance

When the system is calibrated the actual power delivered to the input

"

Reference Plane

" is read on the power meter display for power sensor

A. Sensor A is used to monitor and provide feedback to the user or an

ATE system in order to set the "

Reference Plane

" power level. The

power level is set by adjusting the RF Source power level control.

Calibration of the system is accomplished by removing the DUT and

connecting sensor B to the input "

Reference Plane

" in order to deter-

mine the coupler's coupling coefficient and the other losses to the refer-

ence plane.

These losses are:

PmB - PmA

. These losses are frequency dependant

and must be measured at each test frequency.

Coupler through losses and their coupling coefficients are very linear

over their rated power range and therefore only need to be measured at

one power level.

The frequency dependant losses

(PmB-PmA)

are then provided to the

power meter in the form of a display offset for

PmA

.

The calibration is complete and the DUT input power level is displayed

directly on the power meter sensor A display.

This setup is quick and easy to calibrate and represents a good test

bench for the R&D laboratory, where a manual control system is suffi-

cient to evaluate performance.

Understanding the compression characteristics of the TA2U50HA

System Amplifier is important in these situations and can often be com-

pensated for in the Automated Test Environment (ATE) by using an iter-

ative set and measure routine to set the input reference plane power

level close to the desired level.

Since the incident power is always accurately provided by a

PmA

meas-

urement, the requested incident power level does not always need to be

exact for accurate characterization of a DUT.

TA2U50HA Operating Manual

TA2U50HA Operating Manual

8

5

TA2U50HA Group Delay

Содержание TA2U50HA

Страница 1: ...the test and measurement product portfolio once owned by Centellax Inc is now part of Agilent Technologies For more information about these products and support go to www agilent com find bert news Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 Established 1981 ...

Страница 2: ...centellax com 3843 Brickway Blvd Suite 100 Santa Rosa CA 95403 Tel 707 568 5900 Fax 707 568 7647 Specifications are subject to change without notice All Centellax trademarks and trade names are the property of Centellax Inc Copyright 2003 2009 Centellax Inc Printed in USA 17 Jun 2009 smd 00054 rev B CENTELLAX TA2U50HA ...

Страница 3: ...TA2U50HA Operating Manual TA2U50HA Operating Manual ...

Страница 4: ...rranty is only for the benefice of the original buyer registered with Centellax and is not assignable This warranty does not extend to any of Centellax Products which have been sub ject to misuse neglect accident improper installation or used in violation of operating instruc tions This warranty does not extend to Products which have been repaired calibrated or altered in any way by a facility tha...

Страница 5: ...led to the final output stage of the amplifier The second is a matched diode in close proximity to the detector diode which is not RF coupled but provides a reference voltage that is a function of temper ature and biasing A differential measurement of these diodes provides a voltage proportional to the amplifier output power level independent of temperature and bias level of the output amplifier c...

Страница 6: ... The TA2U50HA features 2 4mm 50GHz female connector at the ampli fier s input and 1 85mm V connector at output V connector is compat ible with 2 4mm Adapters to lower frequency connectors e g 2 9mm 3 5mm and SMA are readily available The TA2U50HA is self contained and plugs into standard AC power sources Rear Panel Input A 2 4 mm input connector is provided on the rear panel of the TA2U50HA Power ...

Страница 7: ...HA Saturated Output Power Typical Module Performance TA2U50HA Operating Manual TA2U50HA Operating Manual 4 9 2 10 GHz 10 30 GHz 30 40 GHz 40 50 GHz Parameter Description Min Typ Max Min Typ Max Min Typ Max Min Typ Max Psat dBm Saturated Output Power 26 30 25 29 23 27 20 24 P 1 dBm 1 dB Compressed Power 24 28 23 27 21 25 17 22 S21 dB Small Signal Gain 25 29 24 28 21 25 17 24 S11 dB Input Match 15 1...

Страница 8: ...ver their rated power range and therefore only need to be measured at one power level The frequency dependant losses PmB PmA are then provided to the power meter in the form of a display offset for PmA The calibration is complete and the DUT input power level is displayed directly on the power meter sensor A display This setup is quick and easy to calibrate and represents a good test bench for the...

Страница 9: ... ured using a power meter with two measurement port sensors A and B e g Agilent E4419B An RF power coupler provides a small proportional amount of the amplified power to power sensor A as shown in Figure 7 Figure 7 Example of an Amplified RF Power Measurement System Section 4 Performance Verification The Figures below show the setup used to verify the performance of the TA2U50HA System Amplifier A...

Отзывы: