•  Determine the line voltage from outlet: 

60 Hz - 208 V or 230 V

   50 Hz - 200 V, 220 V or 240 V

• Select the line voltage.

• Restart system.

10

•  Go to 

www.cem.com/register

 to  

register your system.

• See back of this sheet for details.

12

•  After restart, select “Settings”, then 

“Video”, then “Welcome” for a short 

introduction. (Not available on all 

systems)

11

•  Remove the box from inside of the 

system cavity.

•  Write down the serial number located 

on the back left side of the system.

7

• Lay the sides of the box flat.

•  Remove the foam from the top of  

the system.

4

• Pull plastic tab, swing handle open.

• Remove plastic assembly from box.

•  Repeat process on other 3  

assemblies.

1

•  Push the power button on the side of 

the system to turn it on.

9

•  Remove the plastic and discard.  

Do not leave the plastic under  

the system.

6

•  Carefully cut the tape on both sides 

of the inner box.

3

•  Attach the exhaust hose, and connect 

to back of the system.

•  Plug power cord into the system and 

a dedicated, grounded outlet.

8

•  Lift the system, from the sides, onto 

bench top near exhaust service (two 

people required). Do not place the 

system in a fume hood.  

5

•  Use top handles to pull up and  

remove outer box.

2

Installation Location

1.  Designate an area with a sturdy work-

bench within 8' (2.5 m) of a hood or 

suitable fume disposal mechanism.

2.  Provide at least 8" (20.3 cm) space 

on each side and 3" (7.6 cm) space 

in the rear for ventilation. 

 

System Dimensions:  

21" (53.3 cm) W x 25" (63.5 cm) D x 

25" (63.5) H 

140 lb in weight (without vessels)

 

3.  Allow space for sample handling

4.  The system must be free from the 

vibration of large equipment or  

walk-through traffic

5.  The system must be away from the 

primary exits and walk-through traffic

6.  Location should have a temperature 

of 41°F (5 °C) – 104 °F (40 °C).

7.  The system should be connected to  

a dedicated, grounded outlet.

Ventilation Requirements

The system comes with 8' (2.5 m) of 

flexible exhaust hose (3 

1

8

" OD x 2 

13

16

ID). The built-in exhaust system will push 

fumes through the hose at 125 CFM. 

To properly vent the system, your hood 

or exhaust line must draw at least 125 

CFM at the point of connection.

Electrical Requirements

 

The MARS unit requires 208/230 V, 60 

Hz, 15A @ 230 V, or 200/220/240 V, 50 

Hz, 15A @ 240 V, single phase power. It 

requires a 250 V 15 A rated receptacle 

(NEMA 6-15R) or a 250 V - 20 A rated re-

ceptacle (NEMA 6-20R). It is non-locking. 

MARS Series

Installation Guide