background image

 

www.cem.com        

         [email protected]        

         704-821-7015        

         800-726-3331 

Cleaning and Maintenance Tips for  
Discover® SP-D 10/35 and 80

 

 
 

Discover SP-D Cleaning with Tap Water – to be done DAILY 
 

Note: Complete procedure each day after running samples. The sample will vent 4 to 6 times 
and lose 2 to 3 mL of water  
 

1.

 

Create the following method using the document, CEM Synergy–D Software Operating 
Procedures in the section “Create New Methods.”  

Microwave Power Application: Ramp to Temperature  
Pressure Stages: Custom,  

Stage 1: PressureSP = 125 PSI, Times at SP = 4,  
Stage 2: PressureSP = 250 PSI, Times at SP = 10,  
Remove Stages 3 – 5.  

Vessel: 80 mL Quartz  or 35 mL or 10 mL Pyrex without Liner   
 

Stage  Temp  Ramp 

Time 

Hold 
Time 

Pressure  Power  Stirring 

200 

4:00 

2:00 

400 

300 

Med 

 

2.

 

Add tap water to vessel 

10 mL to the 35 mL and 80 mL vessel, OR  
3 mL and 15 – 20 grains of NaCl to the 10 mL vessel  

3.

 

Add stir bar and cap 

4.

 

Place in Explorer rack or Discover SP–D 

5.

 

Add sample to queue using the document, 

CEM Synergy–D Software Operating 

Procedures

 

in the section “Set-up and Start a Sample Queue with the Explorer” 

OR

 

“Set-up and Start a Sample Queue without the Explorer” 

6.

 

Press Play 

 
 

ActiVent cleaning – to be done WEEKLY 

 

1.

 

Press the HOME key to display the “Discover Home” screen.  

2.

 

Press the “+/-“ keys to highlight “System Setup”.  

3.

 

Press the ENTER key.  

4.

 

Press the right ARROW key 3 times to display the “Setup 4” screen.  

5.

 

Press the “+/-” keys to highlight “APM Cleaning”.  

6.

 

Press the ENTER key. A new screen will appear: 
 

 

 

7.

 

Remove the attenuator (microwave door) from the cavity.  

8.

 

Press the ENTER key. A new screen will appear:  

Sept 2014 

Отзывы: