10 APPENDIX A- Technical references
Cooling curve
Rotor
HF1R
Startup 230Vac / 0.8s / 10A Running 60Vac / 2A
10.5.3
X-Ray source assembly
Model
HF1R
X-ray tube
IAE X22 0.3/0.6
Focus - detector distance
970 mm
Focus - skin distance minimum
150 mm
Total filtration
4.4
mm Al
@
70 kV
Cone beam maximum size
238 mm x 179 mm (detector area)
Reproducibility of the radiation output
Δ < 10%
Tube voltage accuracy
< 10%
Tube current accuracy
< 20%
Radiation linearity
< 20%
Emission time accuracy
< 10% + 1 ms
mAs accuracy
< 30%
8
According to IEC 60601-2-44:2009 par. 203.6.3.2
9
According to IEC 60601-2-7:1998 par. 50.104.1
10
According to IEC 60601-2-7:1998 par. 50.104.2
11
According to IEC 60601-2-7:1998 par. 50.105.4
12
According to IEC 60601-2-7:1998 par. 50.104.3
13
According to IEC 60601-2-7:1998 par. 50.104.3
10-12
NewTom 5G