background image

Instructions 

F20B, F22B

Setup and Preparation

Gauges are calibrated at the factory using equipment traceable 

to NIST. There is no need to calibrate the gauge before putting 

it into service. Calibration should only be performed by qualified 

individuals using appropriate calibration standards and procedures. 

Calibration intervals depend on your quality control program 

requirements, although many customers calibrate annually. 
The calibration system must be able to generate and measure pres-

sure/vacuum over the full range of the gauge and should be at least 

four times more accurate than the gauge being calibrated.
A vacuum pump able to produce a vacuum of 100 microns (0.1 torr 

or 100 millitorr) or lower is required for vacuum gauges. 
Allow the gauge to acclimate to ambient temperature for 20 minutes. 
Install fresh batteries.

Entering Calibration Mode

With the gauge off, press and hold the 

t

 button. Then press the 

Power button. Release all buttons when the display indicates 

CAL

.

The display begins by indicating the full-scale positive pressure 

rating of the gauge in the engineering units as configured by the 

factory, and then shows all display.
Before the gauge enters the Calibration Mode, the display ini-

tially indicates _ _ _ _ with the first underscore blinking, and with 

CALPC

 (calibration pass code) on the lower display. 

Enter the 3510 pass code as described in the Configuration Pass 

Code section.

Calibration Mode

The gauge enters and remains in the Calibration Mode until 

restarted manually or power is removed. Features not related to 

calibration are disabled and compound range models are set for the 

same engineering units for pressure and for vacuum.
The calibration may be performed in any of the available engineer-

ing units as well as percent (

PCT

). 

For greatest accuracy, use the 

s

 and 

t

 buttons to select engi-

neering units for calibration with highest resolution (highest number 

of display counts). 
Press and release the Power button when the appropriate engineer-

ing units are displayed. Suggested units are listed below.
Sensor 

Suggested units for calibration

5 PSI 

5.000 PSI

15 PSI 

775.7 MMHG or TORR

30 PSI 

61.08 INHG

50 PSI 

50.00 PSI

60 PSI 

60.00 PSI

100 PSI 

7.031 KG/CM2

200 PSI 

407.2 INHG

300 PSI 

610.8 INHG

500 PSI 

3447 KPA

1000 PSI 

6895 KPA

2000 PSI 

4613 FTH2O

3000 PSI 

6920 FTH2O

5000 PSI 

5000 PSI

The display will then indicate the currently applied pressure in the 

engineering units selected for calibration.

s

 

and

 

t

 Button Operation

Each time one of the 

s

 or 

t

 buttons is pressed and released 

quickly, a small change is made to the digitized pressure signal. It 

may take more than one of these small changes to result in a single 

digit change on the display. 
To make larger changes, press and hold the appropriate button. 

After about one second, the display will begin to change continu-

ously. Release the button to stop. Then make fine adjustments by 

pressing and quickly releasing the buttons as previously described.

Gauge Reference Pressure Gauges

Apply zero pressure by venting the gauge port to atmosphere. The 

character display will alternate between 

ZERO

 and 

CAL

. Adjust 

for a display indication of zero using the 

s

 and 

t

 buttons.

Apply full-scale pressure. The character display will alternate 

between 

+SPAN

 and 

CAL

. Adjust for a display indication of 

full-scale pressure using the 

s

 and 

t

 buttons. 

Apply 50% full-scale pressure. The character display will alternate 

between 

+MID

 and 

CAL

. Adjust for a display indication equal to 

50% of full-scale pressure using the 

s

 and 

t

 buttons. 

Gauge Reference Vacuum Gauges

Apply zero pressure by venting the gauge port to atmosphere. The 

character display will alternate between 

ZERO

 and 

CAL

. Adjust 

for a display indication of zero using the 

s

 and 

t

 buttons.

Apply full-scale vacuum. The character display will alternate 

between 

+SPAN

 and 

CAL

. Adjust for a display indication of 

full-scale vacuum using the 

s

 and 

t

 buttons. 

Apply 50% full-scale vacuum. The character display will alternate 

between 

+MID

 and 

CAL

. Adjust for a display indication equal to 

50% of full-scale vacuum using the 

s

 and 

t

 buttons. 

Absolute Reference Gauges

Apply full vacuum to the gauge. The character display will alternate 

between 

ZERO

 and 

CAL

. Press the 

s

 and 

t

 buttons to obtain 

a display indication of zero.
Apply full-scale pressure. The character display will alternate 

between 

+SPAN

 and 

CAL

. Press the 

s

 and 

t

 buttons to obtain 

a display indication equal to full-scale pressure.
Apply 50% of full-scale pressure. The lower display will alternate 

between 

+MID

 and 

CAL

. Press the 

s

 and 

t

 buttons to obtain 

an indication equal to 50% of full-scale pressure.

Compound and Bipolar Gauges

In addition to the steps described above for pressure gauges, apply 

full-scale vacuum. The character display will alternate between 

-SPAN

 and 

CAL

. Adjust for a display indication of actual applied 

vacuum using the 

s

 and 

t

 buttons.

For bipolar and –30.15.00psig compound range models 

only, apply 50% full-scale vacuum. The character display will 

alternate between 

-MID

 and 

CAL

. Adjust for a display indication 

equal to 50% of full-scale vacuum using the 

s

 and 

t

 buttons. 

Save Calibration

Press and hold the Power button until the display indicates 

- - - -

 

then release the button to store the calibration parameters in non-

volatile memory and restart the gauge.
Verify the pressure indications at 0%, 25%, 50%, 75% and 100% 

of full scale.

The factory default pass code 3510 may be changed to a different 

value for configuration and/or calibration.

Configuration Pass Code

With the unit off, press and hold the 

s

 button to view and/or 

change the user configuration pass code. Then press the Power 

button. Release all buttons when the display indicates 

CFG

.

Calibration Pass Code

With the unit off, press and hold the 

t

 button to view and/or 

change the user calibration pass code. Then press the Power but-

ton. Release all buttons when the display indicates 

CAL

.

Change Pass Code Mode

Before the unit enters the view or change pass code mode, the 

display initially indicates _ _ _ _ with the first underscore blinking, 

and with 

CFGPC

 or 

CALPC

 on the character segments. 

Note: The unit will automatically revert to normal operation if no 

buttons are operated for approximately 15 seconds. To cancel and 

return to normal operation, press and release the Power button 

without entering any pass code characters.
Enter access code 1220:
Use the 

s

 and 

t

 buttons to set the left-most digit to 1.

Press and release the Power button to index to the next position. 

The 1 will remain, and the second position will be blinking.
Use the 

s

 and 

t

 buttons to select 2.

Press and release the Power button to index to the next position. 1 

2 will remain, and the third position will be blinking.
Use the 

s

 and 

t

 buttons to select 2.

Press and release the Power button to index to the next position. 1 

2 2 will remain, and the fourth position will be blinking.
Use the 

s

 and 

t

 buttons to select 0.

Press and release the Power button to proceed. 
Note: If an incorrect access code was entered, the gauge will return 

to the start of the access code entry sequence.

Change Pass Code

Once the access code has been entered correctly, the display will 

indicate the existing user-defined pass code with 

CFGPC

 or 

CALPC

 on the character segments.

Press the 

s

 or 

t

 button to select the first character of the new 

pass code.
When the correct first character is being displayed, press and 

release the Power button to proceed to the next pass code character.
Repeat above until the entire pass code is complete.
To exit the User Defined Pass Code change mode, press and hold 

the Power button. 
Release the button when the display indicates 

- - - -

  to restart 

the gauge.

Calibration

User-Defined Pass Code Configuration

Calibration

continued

Battery Replacement

A low battery indication will be shown in the upper left-

hand corner of the display when the battery voltage falls 

sufficiently. The batteries should be replaced soon after 

the indicator comes on or unreliable readings may result.
1.  Remove the 6 Phillips screws on the back of the unit. 
2.  Lift up the battery holder.
3.  Remove batteries by lifting up the positive end of the battery 

(opposite the spring) taking care not to bend the battery holder 

spring.

4.  Discard old batteries properly, do not discard into fire, sources 

of extreme heat, or in any hazardous manner.

5.  Always replace both batteries at the same time with high quality 

alkaline batteries.

6.  Install batteries with correct orientation. Incorrect polarity will 

damage the gauge. The negative (flat) end of each battery 

should be inserted first facing the battery holder spring.

7.  Replace battery holder and back cover, including the rubber 

gasket and reinstall the six screws.

The M8 version allows recording of up to eight pressure readings. 

While in the memory mode the peak reading is captured.
The eight MEM 1 through MEM 8 labels are factory set. 

F20B Memory

M8 Versions

F20B Memory

M4 Versions

The M4 version allows recording pressure 

readings of up to four tires. While in the 

memory mode the peak reading is captured.
The number 1 is shown on the upper dis-

play. The lower display will indicate the label 

for memory 1.
Use the 

s

 and 

t

 buttons to select the 

desired label: 

MEM I

LR

 (left rear), 

RR

 

(right rear), 

RF

 (right front), or 

LF

 (left front). 

Each of the memory locations may be 

renamed as desired in any sequence. Care 

should be taken to avoid duplication or 

omission of a position.
When the desired label for memory 1 is 

displayed, press the Power button.
Repeat the steps for the other memory locations. When the desired 

label for memory 4 is displayed, press and release the Power button to 

save the user configuration and restart the gauge.

I

Mem I

- or -

I

  LF 

t

s

Zero/Clear

Memory

t

s

Zero/Clear

Memory

*

*

*

If a special memory configuration was ordered, 

the custom memory labels may be shown. 
In some cases, it may be possible to rename 

the MEM locations to the custom labels. If this 

is the case, use the following example below.
After auto shutoff time selection, the number 

1 is on the upper display. The lower display 

will indicate the label for memory 1. 
Use the 

s

 and 

t

 buttons to select MEM 1, 

or the programmed memory label.
When the desired label for memory 1 is dis-

played, press and release the Power button.
Repeat the steps for the other memory 

locations. When the desired label for the last 

memory location is displayed, press and release the Power button 

to save the user configuration and restart the gauge.

F20B Memory

Custom Versions

I

Mem I

I

 NLG I 

t

s

Zero/Clear

Memory

*

*

- or -

t

s

Zero/Clear

Memory

*

F22B Backlight Time Selection

The lower display will indicate 

BL 

if the 

display backlight is enabled or 

NO BL

 if 

display backlight is disabled. 
Use the 

s

 button to enable backlighting 

and the 

t

 button to disable backlighting.

Press the power button to save the setting.
If 

NO BL

 was selected the user setup is 

complete and the gauge will restart and be 

ready for use with the new configuration. 
If

 BL

 was selected the current backlight 

auto shutoff time is displayed in minutes.  

1 minute is the factory default. 
Use the 

s

 and 

t

 buttons to select the 

minutes for backlight shutoff time.
A setting of zero disables the auto shutoff timer and the backlight 

will be on whenever the gauge is on. The maximum setting is 255 

minutes. The gauge auto shutoff time will override the backlight 

time.  When the desired time is displayed, press and release the 

power button to save your selection and restart the gauge.

t

s

Zero/Clear

Memory

*

*

BL

- or -

No BL

t

s

Zero/Clear

Memory

2

BLTMR

*

*

- or -

cecomp.com

1220 American Way Libertyville, IL 60048 

800-942-0315

CECOMP

 

Div. of

 

A

BSOLUTE

 P

ROCESS

 I

NSTRUMENTS

Содержание F20B

Страница 1: ...1 35KGCMG 01 350GCMG 1 0 1MPAG 0001 2ATMG 001 7000GCMG 1 35ATMG 01 260MMHGG 1 1BARG 001 60 psig Res 5200MMHGG 1 1000 psig Res 260TORRG 1 1KGCMG 001 60PSIG 01 5200TORRG 1 1000PSIG 1 350MBARG 1 1ATMG 001 120INHGG 1 7000MBARG 1 2040INHGG 1 350CMH2OG 1 15 psig Res 1660INH2OG 1 7000CMH2OG 1 2300FTH2O 1 3500MMH2OG 1 15PSIG 01 960ZING 1 230FTH2O 1 7000KPAG 1 12FTH2O 01 30INHG 15PSIG 01 4200GCMG 1 700KPAG...

Страница 2: ... 8 2068 2041 2109 210 9 2109 210 9 48 00 6 921 83 0 155 1 155 1 6 108 0 to 5 000 psig 5 000 34 47 344 7 3447 3402 3515 351 5 3515 351 5 80 0 11 54 138 4 258 6 258 6 10 18 15 00 to 0 psi abs 15 00 abs 103 4 abs 1034 abs 1034 abs 1 034 abs 1 021 abs 1 055 abs 1055 abs 1055 abs 240 0 abs 34 61 abs 415 1 abs 775 7 abs 775 7 abs 30 54 abs 0 to 14 70 psig vac 14 70 vac 101 3 vac 1013 vac 1013 vac 1 013 ...

Страница 3: ...ress and release the Memory button to view the maximum stored value The center power button may be pressed at any time to return to the normal display mode The gauge may be left in the maximum dis play mode if desired The maximum reading will be continuously displayed stored and updated Press and release the Memory button to view the minimum stored value For many applications it may be best to bri...

Страница 4: ...ation Pass Code With the unit off press and hold the t button to view and or change the user calibration pass code Then press the Power but ton Release all buttons when the display indicates CAL Change Pass Code Mode Before the unit enters the view or change pass code mode the display initially indicates _ _ _ _ with the first underscore blinking and with CFGPC or CALPC on the character segments N...

Отзывы: