background image

cecomp.com

1220 American Way Libertyville, IL 60048
Phone: 

800-942-0315

 

Fax: 800-949-7502

Instructions F16AD, 

F16ADN

BSOLUTE

ROCESS

NSTRUMENTS, Inc.

Division of

  Read these instructions before using the gauge. Configuration 

may be easier before installation. Contact the factory for assis-
tance.

  These products do not contain user-serviceable parts. Contact 

us for repairs, service, or refurbishment.

  Gauges must be operated within specified ambient temperature 

ranges.

  Outdoor or wash down applications require a NEMA 4X gauge or 

installation in a NEMA 4X housing.

  Use a pressure or vacuum range appropriate for the application.

  Use fittings appropriate for the pressure range of the gauge. 

  Due to the hardness of 316 stainless steel, it is recommended 

that a thread sealant be used to ensure leak-free operation.

  For contaminated media use an appropriate screen or filter to 

keep debris out of gauge port. 

  Remove system pressures before removing or installing gauge.

  Install or remove gauge using a wrench on the hex fitting only. 

Do not attempt to turn gauge by forcing the housing. 

  Good design practice dictates that positive displacement liquid 

pumps include protection devices to prevent sensor damage 
from pressure spikes, acceleration head, and vacuum extremes.

 

Avoid permanent sensor damage! Do not apply vacuum to non-
vacuum gauges or hydraulic vacuum to any gauges. 

 

Avoid permanent sensor damage! NEVER insert objects into 
gauge port or blow out with compressed air.

 Gauges are not for oxygen service. Accidental rupture of sensor 

diaphragm may cause silicone oil inside sensor to react with 
oxygen.

Bipolar ranges read positive pressure and vacuum in the same 
units, and zero with the gauge port open.

1000 psi and higher sensor are a sealed reference type. They  read 
zero with the gauge port open are internally referenced to 14.7 psi. 
Functionally similar to gauge reference sensors.

Absolute reference gauges read zero at full vacuum and atmo-
spheric pressure with the gauge port open. With an open gauge 
port the readings will vary continuously due to the effects of 
barometric pressure. 

Range Codes

The range code is part of the gauge model number and indicates 
the default range when the gauge is ordered. 

Selectable Ranges

Engineering units may be changed to any of those listed in the same 
sensor range as shown in the range table. 

Conversion

Engineering units are calculated from the factory default unit to the 
newly selected units. The ranges listed under the psi sensor range 
are rounded off.

The AD series is powered by 8-24 VAC 50/60 Hz or 9-32 VDC.

The type and magnitude of the supply voltage have negligible 
effects on the gauge calibration as long as it is within the voltage 
ranges stated above. No polarity needs to be observed when con-
necting a power supply. An inexpensive unregulated low voltage AC 
or DC power supply can be used.

After the gauge is installed, route the wires away from heat sources 
and moving equipment and connect the low-voltage power source 
to the gauge wires.

Ensure that the gauge supply voltage does not fall below 8 VAC

RMS

 

if AC power is used, or 9 VDC if DC power is used. Operation with 
less than these values may cause erratic or erroneous readings. 

When operating multiple gauges from the same power supply, refer 
to the mA rating in the specifications to ensure adequate power.

Note that standard 24 VAC transformers with small loads often 
operate at voltages well over the gauge's 24 VAC limit. 

Power-Up

Press and hold the front button for approximately 1 second.

The display is tested, the default full-scale range is indicated, the 
full-scale range in the selected units are indicated, the display is 
tested again, then the actual pressure and units are displayed.

Power-Up With Zero

This applies to gauge reference models only. Absolute reference 
gauges do not use the zero feature since they read atmospheric 
pressure under normal conditions.

Be sure the gauge port is exposed to normal atmospheric pressure 
and no pressure is applied. The zeroing function is only activated 
at each power-up and the stored zero correction is erased when 
the gauge is shut off.

Press and hold the front button. The display is tested and then oooo 
is displayed. The gauge is now zeroed. Release the button and the 
full-scale range in the selected units are indicated, the display is 
tested again, then the actual pressure and units are displayed.

Attempting to zero the gauge with pressure greater than approxi-
mately 3% of full-scale pressure or vacuum applied will result in an 
error condition, and the display will alternately indicate Err 0 and 
the actual measured pressure. The gauge must be powered down 
to reset the error condition. 

Normal Operation

Following the start-up initialization, the display indicates the pres-
sure reading updated approximately 3 times per second. The gauge 
may be left on continuously or shut off as desired. 

If excessive vacuum is applied to a pressure-only gauge, the display 
will indicate -Err until the vacuum is released. 

Applying vacuum to a gauge designed for pressure may damage 
the pressure sensor. If excessive pressure is applied (112.5% over 
range), an out-of-range indication of 1 – – – or 1.–.–.– will be 
displayed depending on model.

Display Backlighting (BL models only)

Display backlighting is on whenever the gauge is on. 

The display backlighting will not be apparent under bright lighting 
conditions.

Minimum and Maximum Readings

Gauges are normally configured with minimum and maximum 
capture functions enabled. One or both can be disabled in the User 
Configuration mode. 

Minimum and maximum readings are continuously stored and 
updated whenever the gauge is on. The stored readings can be 
manually cleared if desired. The MAX and MIN memory is also 
cleared whenever the gauge is off unless configured to save the 
readings.

Press and hold the button for about 1 second until MAX is displayed 
alternating with the units. The maximum reading will be continu-
ously updated. The gauge may be left in this mode.

After MAX is displayed, press and hold the button for about 1 sec-
ond until MIN is displayed alternating with the units. The minimum 
reading will be continuously updated. The gauge may be left in this 
mode. If excessive vacuum is applied to a pressure-only gauge 
while in this mode, the display will indicate -Err until the MAX/MIN 
readings are cleared. 

After MIN is displayed, press and hold the button again for about 1 
second until * * * * is displayed. The MAX and MIN memory is not 
erased and the gauge returns to normal operation with the display 
indicating the current reading.

Press and continue to hold the button until the display indicates clr 
MX/MN (about 3 seconds total) and then release the button. Both 
maximum and minimum values are cleared and the gauge returns 
to the normal mode and displays the current pressure.

Shut-Down

To shut off the gauge manually at any time, press and hold the button 
until the display indicates OFF (about 5 seconds) and then release.

Operation—continued

Types of Gauges

Installation Precautions

Dimensions

Operation

Cecomp maintains a constant effort to upgrade and improve its products. 
Specifications are subject to change without notice. Consult factory for your 
specific requirements.

Selectable Ranges

¼" NPT

2.88"

3.38" Standard

3.5" NEMA 4X

0.75"

2.0"

Cecomp Electronics / Made in USA

1.65"

NEMA 4X

3.0"

Standard

Turn at hex

fitting only

X X X X X

88888

2

Low voltage 

indicator

3 ft long 2 
conductor 22 
AWG cable with 
stripped and 
tinned wire 
ends

Turn at hex

fitting only

Function

Button

Prompt (Release Button)

On

Press for 1 

second

Display test > default units > selected 
units 

FUll scale 

> display test > 

actual reading

Zero

(gauge ref. 

only)

Press and 

hold

Display test > 

oooo

 > default units > 

selected units 

FUll scale 

display test > actual reading

Hi reading

Press/hold

Max 

> max. reading & units

Lo reading Press/hold

MIN 

> min. reading & units

Exit Hi/Lo

Press/hold



 

> actual reading

Clear Hi/Lo  Press/hold

Max 

Clr mx/mn

 > actual reading

Clear, off

Press/hold

Max 

Clr mx/mn

 > 

OFF

Power

MOD-TRONIC Instruments 

·

 

Brampton, ON 

· 

905-457-6322 or 1-800-794-5883 

·

 

[email protected] 

· 

www.mod-tronic.com

 

Отзывы: