C. Unit Selection
1. Press the UNIT button to select the desired unit.
The display switches between
lb:oz
and
kg.g
.
2.
oz
appears on the display when the weight is
less than 1 lb. Otherwise, both
lb
and
oz
appear
on the display.
3.
g
appears on the display when the weight is less
than 1 kg. Otherwise,
kg
appears on the display.
D. Hold Mode
1. Press and hold the HOLD button for 2 seconds.
HOLD
flashes on the display for 15 seconds to
allow enough time to place an item on the scale.
Note:
If no item is placed on the scale, the scale will exit
Hold Mode and return to Weighing Mode.
2. When the item is placed on the scale, the
item’s weight and
HOLD
appear on the display
for 15 seconds.
3. When the hold time is over, the scale exits
Hold Mode automatically and returns to
Weighing Mode.
Note:
During Hold Mode all buttons are locked.
E. Weighing Mode
1. Press the ON/OFF button to turn on. The display
flashes
–––
and then returns to zero.
2. Place an item on the scale platform. The weight
of the item appears on the display within
1 second after the item is stable.
Note:
The minimum measurable weight is 0.02 oz/0.5 g.
F. Tare
The tare function resets the scale display to zero
in order to measure only the weight of an item.
It is frequently used to subtract the weight of
containers.
1. When the scale is on, place a container on
the platform.
2. Press the TARE button to reset the display
to zero. The display flashes
–––
and then returns
to zero.
3. Place the item into the container or directly on
the scale platform. The weight of only the last
item added will appear on the display.
4. Continue to repeat steps
2.
and
3.
to measure
the respective weights of additional items.
5. To clear the tare, remove all items from the
platform and press the TARE button. The display
is reset to zero.
G. Field Calibration
1. When scale is off, press and hold the ON/OFF
button until
–––
is displayed.
2. Press the UNIT button 3 times within two
seconds.
3. Press and hold the UNIT button until
0
appears
on the display. The 5-digit internal code value
then appears on the display.
4. Press the UNIT button again. The software
version and then
1000.0 g
appears on
the display.
Note:
To change the calibration units to grams or pounds,
press the UNIT button. 1 kg (WT01) calibration weights are
available from CDN.
5. Put a 1 kg calibration weight in the center of the
platform. The display becomes blank briefly and
then
PASS
appears on the display for 3 seconds
if the calibration is successful. The scale has
been calibrated and is ready for use.
H. Battery Status
bAtt
:
replace battery with two AAA alkaline cells.
The scale turns off automatically after 2 seconds.
I. Error Messages
1.
EEEE
indicates an overload when the item
being measured exceeds the capacity of the
scale. Immediately remove the overweight item
to avoid causing permanent damage to your
scale. The scale turns off automatically after
2 seconds.
2.
Err2
indicates an initialization problem.
The scale turns off automatically after
2 seconds. Turn on the scale again. If the
problem persists, please contact CDN.
Note:
Clean the scale platform before each use. The
stainless steel removable platform is dishwasher safe; the
removable plastic platform base is not dishwasher safe.
Note:
Replacement removable platform, AC adapter and
1 kg calibration weights available.
CE Note:
This device could be sensitive to electrostatic
discharge. If electrostatic discharge or malfunctioning
occurs, please re-install the battery to reset this unit.