C. Unit Selection

1.  Press the UNIT button to select the desired unit. 

The display switches between 

lb:oz

 and 

kg.g

.

2. 

oz

 appears on the display when the weight is 

less than 1 lb. Otherwise, both 

lb

 and 

oz

 appear 

on the display.

3. 

g

 appears on the display when the weight is less 

than 1 kg. Otherwise, 

kg

 appears on the display.

D. Hold Mode

1.  Press and hold the HOLD button for 2 seconds. 

HOLD

 flashes on the display for 15 seconds to  

allow enough time to place an item on the scale. 

Note:

 If no item is placed on the scale, the scale will exit 

Hold Mode and return to Weighing Mode.

2. When the item is placed on the scale, the  

item’s weight and 

HOLD

 appear on the display 

for 15 seconds.

3. When the hold time is over, the scale exits  

Hold Mode automatically and returns to  
Weighing Mode. 

Note:

 During Hold Mode all buttons are locked.

E. Weighing Mode

1.  Press the ON/OFF button to turn on. The display 

flashes 

–––

 and then returns to zero. 

2. Place an item on the scale platform. The weight 

of the item appears on the display within  
1 second after the item is stable. 

Note:

 The minimum measurable weight is 0.02 oz/0.5 g.

F. Tare 

The tare function resets the scale display to zero 
in order to measure only the weight of an item. 
It is frequently used to subtract the weight of 
containers.
1.  When the scale is on, place a container on  

the platform. 

2. Press the TARE button to reset the display  

to zero. The display flashes 

–––

 and then returns 

to zero.

3. Place the item into the container or directly on 

the scale platform. The weight of only the last 
item added will appear on the display.

4. Continue to repeat steps 

2.

 and 

3.

 to measure 

the respective weights of additional items. 

5. To clear the tare, remove all items from the 

platform and press the TARE button. The display 
is reset to zero.

G. Field  Calibration

1.  When scale is off, press and hold the ON/OFF 

button until 

–––

 is displayed.

2. Press the UNIT button 3 times within two 

seconds. 

3. Press and hold the UNIT button until 

0

 appears 

on the display. The 5-digit internal code value 
then appears on the display.

4. Press the UNIT button again. The software 

version and then 

1000.0 g

 appears on  

the display. 

Note:

 To change the calibration units to grams or pounds, 

press the UNIT button. 1 kg (WT01) calibration weights are 

available from CDN.

5. Put a 1 kg calibration weight in the center of the 

platform. The display becomes blank briefly and 
then 

PASS

 appears on the display for 3 seconds 

if the calibration is successful. The scale has 
been calibrated and is ready for use.

H. Battery  Status 

bAtt

:

 replace battery with two AAA alkaline cells.  

The scale turns off automatically after 2 seconds.

I.  Error Messages

1. 

EEEE

 indicates an overload when the item  

being measured exceeds the capacity of the 
scale. Immediately remove the overweight item 
to avoid causing permanent damage to your 
scale. The scale turns off automatically after  
2 seconds.

2. 

Err2

 indicates an initialization problem.  

The scale turns off automatically after  
2 seconds. Turn on the scale again. If the 
problem persists, please contact CDN.

Note:

 Clean the scale platform before each use. The 

stainless steel removable platform is dishwasher safe; the 

removable plastic platform base is not dishwasher safe. 

Note:

 Replacement removable platform, AC adapter and  

1 kg calibration weights available.

CE Note:

 This device could be sensitive to electrostatic 

discharge. If electrostatic discharge or malfunctioning 

occurs, please re-install the battery to reset this unit.

Содержание SD0204

Страница 1: ... 1 Power off the unit before installing the batteries 2 Remove battery cover on the bottom by pressing the latch toward the battery cover and lifting up 3 Install two AAA alkaline batteries observing polarity shown in compartment 4 Replace the battery cover until it clicks shut POWER ADAPTER Before using the external power adapter make sure that the printed voltage complies with the local voltage ...

Страница 2: ...ctive weights of additional items 5 To clear the tare remove all items from the platform and press the TARE button The display is reset to zero G Field Calibration 1 When scale is off press and hold the ON OFF button until is displayed 2 Press the UNIT button 3 times within two seconds 3 Press and hold the UNIT button until 0 appears on the display The 5 digit internal code value then appears on t...

Страница 3: ...h abrasive or corrosive compounds which may scratch the surface and corrode the electronic circuits Do not subject the scale to excessive force shock dust temperature or humidity which may result in malfunction shorter electronic life span damaged battery and or distorted parts Do not tamper with the scale s internal components Doing so will invalidate the warranty and may cause unnecessary batter...

Страница 4: ...isen by using this product or resulting from any defect omission in this document even if advised of the possibility of such damages The manufacturer and its affiliates reserve the right to make improvements or changes to this document and the products and services described at any time without notice or obligation 1 Year Limited Warranty Any instrument that proves to be defective in material or w...

Отзывы: