3. USAGE WARNINGS
3.1
Allowed Usage
D
D
The instrument has been projected as measure and
control device, built according to EN61812-1 for the
altitudes operation below 2000 ms.
Using the instrument for applications not expressly permitted
by the above mentioned rule must adopt all the necessary
protective measures.
The instrument
must not be used in dangerous environ-
ments
(flammable or explosive) without adequate protections.
D
D
The installer must ensure that the EMC rules are
respected, also after the instrument installation, if
necessary using proper filters.
4. INSTALLATION WARNINGS
4.1
Mechanical Mounting
The instrument, in 78 x 35 mm case, is designed for flush-
in panel mounting. Make a 71 x 29 mm hole and insert the
instrument, fixing it with the provided special brackets.
To obtain the declared protection degree (IP65), use screw
type bracket (optional).
Avoid placing the instrument in dirty environments or with very
high humidity levels that may create condensation and avoid
the introduction of conductive substances into the instrument.
Ensure adequate ventilation to the instrument and avoid
installation in containers that house devices which may
overheat or which may cause the instrument to function at a
temperature higher than the one permitted and declared.
Connect the instrument as far away as possible from sourc-
es of electromagnetic disturbances such as motors, power
relays, relays, solenoid valves, etc..
4.2
Mechanical Dimensions
[mm]
4.2.1
Instrument dimensions
P
U
CNT
Start / Stop
OUT1
OUT2
1
2
35
78
12.2
14.5
8
6
64
14.5
28
4.2.2
Panel cutout
Recommended
panel cutout
29
71
min. 12 mm
min.
12 mm
4.2.3
Mounting brackets
12 mm max.
P
anel + Gask
et
“Butterfly” type brackets
“Screw type” bracket
34
ELETTROMECCANICA CDC S.r.l - TX40
- MANUAL rev.2 - PAG.
4
TX40